3-10
Describing the MCS+ Control Panel
P/N 85213-30, Manual revision: B
Yes +/No – program keys
To modify a displayed parameter value as it appears on the MCS+ display screen,
the operator should:
"
Press Yes + to increase the value.
"
Press No – to decrease the value.
The parameters will be altered according to specific increments as determined by
Haemonetics. Default values and increments for modification will be listed in
each respective protocol manual.
Help/Haemo Calculator key
This key is used during MCS+ operation for different purposes, depending on the
operating state of the device.
During normal MCS+ protocol operation, the operator can:
"
Consult the Haemo Update display screen by pressing this key once.
"
Consult the Haemo Calculator display screen by pressing this key twice.
If the operator receives a NOTICE display screen due to an error detection by the
MCS+ safety system, the operator can press this key to receive the HELP screen
display.
The HELP screen will contain the most likely source of error and a list of the
appropriate operator actions.
Note: A complete listing of all MCS+ operating NOTICE messages and related
HELP messages is provided in the Section 1 of the Postscript to the MCS+.
AC Volume Used
28 ml
Elapsed Time
0 min
Target Cycles
8
NaCl Volume Used
0 ml
HAEMO UPDATE
Plasma Weight
0 g
Platelet Weight
0 g
Estimated Yield
0.0 10e11
Target Yield
5.0 10e11
LDP
STOP
Cycle 0
0 Platelet Weight
0 ml Processed
Press STOP to return to Main display.
Press HELP for the Haemo Calculator.
HAEMO CALCULATOR
Sex
M
Height
170 cm
Weight
65 kg
Blood Volume
4800 ml
HCT
40 %
Plt Pre-Count
250 10e3
Target Plasma Wgt
0 g
Target Yield
5.0 10e11
*Process Volume
4000 ml
Target Cycles
8
Time (estimated)
80 min.
LDP
STOP
Cycle 0
0 Platelet Weight
0 ml Processed
Press MODIFY to select, +/- to change values.
Press SAVE to save all values.
Press HELP to return to Main display.
Содержание MCS+
Страница 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Страница 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Страница 20: ......
Страница 30: ......
Страница 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Страница 85: ......
Страница 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Страница 100: ......