This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
TRN-QRG-100125-US(AC)
Disconnect and Clean Up
When procedure is complete:
1. Touch
End Procedure
, then confirm by touching again
End Procedure
. System empties blue line and
Procedure Complete
screen appears.
2.
Purge extra air from RBC bag (optional):
Touch
remove air
from the procedure complete screen and
follow screen instructions.
3. Close all clamps on all disposable tubing.
4. Cap all ports on collection reservoir.
NOTE:
Tubing stubs and caps included with reservoir and processing set may be used for closing open ports.
5. Remove waste bag from support pins (may be emptied before removing).
6. Remove bowl from centrifuge well.
7. Remove remaining tubing harness.
8. Remove collection reservoir; keep tubing harness connected to reservoir outlet to avoid fluid spill.
9. Discard Cell Saver Elite+ disposables/waste according to the facility’s policies and procedures for handling
of biohazardous materials.