TRN-QRG-100125-US(AC)
This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
Emergency Options (page 2 of 3)
Emergency Mode
During Emergency Mode the device processes blood continuously at high speeds through the Fill, Wash, and Empty phases
until the air detector senses air for the first time in the Fill phase, indicating that the reservoir is empty. The device then reverts
to the previous settings group and enters Standby.
NOTE:
Emergency Mode is not available with the 70 mL bowl.
NOTE:
The line sensor is not active during Emergency Mode.
To enter Emergency Mode:
1. Touch
Emergency Mode
.
2. Touch
On
to confirm.
Pump speed increases to process faster.
RBCs may spill into waste bag during
Emergency Mode.
Device will continue FILL-WASH-EMPTY until no more
RBCs remain in reservoir.
NOTE:
Once the reservoir is empty it will return to automatic mode.
When it starts processing again, Emergency Mode will be
turned off.
To exit Emergency Mode before reservoir
is empty:
1. Touch
Emergency Mode
.
2. Touch
Off
to confirm.