Connectivity

P/N 120857-US, Manual Revision: AC

Haemonetics

®

 Cell Saver

® 

Elite

®

/Elite

®

+ User Manual

To generate a network configuration QR code:

1. Using a computer, download the CSEliteNetConfig utility from the 

Haemonetics Software Download Center.

a. Go to 

http://www.haemonetics.com

, select 

LOGIN

, and click on 

Software Download Center

 in the drop-down menu.

Figure 2, Accessing the Software Download Center

b. Enter your customer login credentials.

Note: If you do not already have a login, contact the Haemonetics 
Customer Care Center to request access.

c. Locate the CSEliteNetConfig Utility within the designated Cell Saver 

Elite+ product area. 

d. Click on 

CSEliteNetConfig Utility

.

e. Download the html file.

2. Open 

CSEliteNetConfig.html

 and ensure that javascript is enabled.

3. Complete the form with the appropriate network settings and click 

Generate QR Code

. The QR code appears.

Table 1, Options for Wireless Authentication

Value

Description

TLS

802.1X/WPA-Enterprise/WPA2-Enterprise

EAP-FAST

802.1X/WPA-Enterprise/WPA2-Enterprise

TTLS

802.1X/WPA-Enterprise/WPA2-Enterprise

LEAP

802.1X/WPA-Enterprise/WPA2-Enterprise

PEAP

802.1X/WPA-Enterprise/WPA2-Enterprise

PSK

WPA-Personal/WPA2-Personal

Содержание Cell Saver Elite

Страница 1: ...Connectivity Upgrade User Manual Addendum 120857 US AC June 2017 0123...

Страница 2: ...tain Haemonetics Corporation products This manual should be used in conjunction with instruction and training supplied by qualified Haemonetics personnel Any failure to follow the instructions as desc...

Страница 3: ...onetics Corporation in the United States and or other countries Bluetooth is a registered trademark of Bluetooth SIG Inc QR Code is a registered trademark of Denso Wave Inc DELL and OptiPlex are trade...

Страница 4: ......

Страница 5: ...ctivity 7 Radio Equipment Specifications 7 Wired Connectivity 7 Network Setup and Maintenance 9 Wireless Network Setup and Maintenance 9 Creating a Network Configuration QR Code 9 Configuring the Devi...

Страница 6: ...the end of the procedure The device must receive an acknowledgment from the server application before it marks the data as sent Status update messages The device sends messages to the server applicat...

Страница 7: ...ol Wireless Connectivity The Cell Saver Elite Elite device is equipped with an internal wireless antenna and wireless module communication component that enable the device to communicate with an appro...

Страница 8: ...sion AC Haemonetics Cell Saver Elite Elite User Manual Figure 1 Parts of the device display 2 6 4 1 3 5 7 1 Status beacon 2 Touch screen 3 Barcode reader laser aperture 4 STOP key 5 Touch screen mount...

Страница 9: ...ss Relocate equipment change channels or follow wireless equipment manufacturer instructions to reduce or eliminate interference Cell Saver Elite Elite devices should be kept a minimum distance away f...

Страница 10: ...e Haemonetics Customer Care Center to request access c Locate the CSEliteNetConfig Utility within the designated Cell Saver Elite product area d Click on CSEliteNetConfig Utility e Download the html f...

Страница 11: ...ustomer Care Center by calling 800 537 2802 or by emailing productsupport haemonetics com Configuring the Device The Network Settings screen is password protected and you must have administrator acces...

Страница 12: ...d by Network Connected and then Server Connected once it is connected to the server application 9 Touch Done to return to the System screen Note In the System screen you can also export or import netw...

Страница 13: ...ce data transmission in the network using the Haemonetics HaemoCommunicator software as the data collection application The list is for reference only and is not intended to be a list of required equi...

Страница 14: ...occurs between connected devices on the wireless network power off the router wireless access point for 30 seconds and then power it back on Parameter Value Model D LINK DIR 803 Channels Four 10 100 E...

Страница 15: ...AC Ensure the wireless network router s provide enough capacity to connect to the server Haemonetics devices and other devices that may run at the same time Each of these will use a connection on the...

Страница 16: ...Once the data is marked as sent the procedure record displays the date and time of export If the Cell Saver Elite Elite device is not connected to the server application the device automatically send...

Страница 17: ...record will not be sent to the server application at the end of the procedure In this situation the Procedure Complete message and the status beacon will be yellow instead of green If you access the R...

Страница 18: ...ng determining which fields appear in the Records screen and making a field mandatory Note You must have administrator level access in order to unlock the Options screen To make a field mandatory 1 To...

Страница 19: ...hich will take several minutes Corrective Action Touch Install Update to install the software now or Touch Skip Update to skip the software installation This prompt will recur on the next power up 503...

Страница 20: ......

Отзывы: