BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
12
•
Do not reuse, recharge, or reheat an old battery cell.
•
Do not attempt to force open the battery cell.
•
Do not discard used batteries with regular municipal/household waste
—see
Appendix C.2
—Disposing of Electrical/Electronic Equipment
•
Discard used batteries according with local waste regulations
—for example,
WEEE, EU battery directive).
The battery located within the computer module of the BloodTrack Kiosk V5 should only
be replaced by Haemonetics personnel as required during part of servicing routine.
Under normal operation where the BloodTrack Kiosk V5 remains powered, 24/7 battery
replacement will be a time period measured in the lifetime of the battery, which is
approximately 10 years.
The replacement battery should match the following specifications:
Vic-dawn
Enterprise KTS
CR2032
BCR2032H13.0AM1UB
3.0V,
220mAh
Compliant
with UL 1642
Compliant
with IEC
80086-4
UL component
file (MH20550)
2.11 Relay Output Specification
The relay output is only for switching less than or equal to a 48V DC or 48V AC RMS
signal up to 1A current.
If higher current, voltage or mains powered switching is required, contact Haemonetics
Technical Support for advice.
2.12 Maintenance Specification
CAUTIONS
•
Only qualified personnel (from the responsible healthcare site, Haemonetics
representatives, or Haemonetics-approved representatives), should perform
the Portable Appliance Testing (PAT) on the BloodTrack Kiosk V5.
•
Only Haemonetics-approved service personnel and/or
Haemonetics-approved representatives should perform any service or
maintenance to the BloodTrack Kiosk V5.
•
The BloodTrack Kiosk V5 should only be used with the cordsets listed in
Section 2.6
•
Haemonetics is not responsible for regulatory compliance of a modified
Haemonetics product.
The BloodTrack Kiosk V5 has no user-serviceable parts.
Содержание BloodTrack Kiosk V5
Страница 2: ...2016 Haemonetics Corporation...