BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
6
2.4
Environmental Specifications
The following environmental conditions should be respected pertaining to operation and
storage of the BloodTrack Kiosk V5:
Conditions
Values
Ambient operating temperature
32° F to 104° F
0° C to 40° C
Storage/transportation temperature
-14° F to 158° F
-10° C to 70° C
Relative humidity RH
10% ~ 90% @ 45°C, non-condensing
Atmospheric pressure range
Maximum a.s.I. 6562 ft.
2000 m
Vibration Loading During Operation
5Grms, IEC 60068-2-64, random, 5 ~ 500Hz, 1
Oct./min, 1hr/axis
Shock During Operation
50G, IEC 60068-2-27, half size, 11ms duration
2.5
Electrical Specifications
CAUTIONS
•
When replacing the fuse for the mains supply, use
only
the fuse referenced
below.
•
The BloodTrack Kiosk V5 utilizes double pole live/neutral fusing.
The electrical specifications for operating the BloodTrack Kiosk V5 are:
Item
Description
AC input
100-240 VAC
Current
1.3 A (rms) for 100 VAC
0.7 A (rms) for 240 VAC
Frequency
47-63 Hz
Earth leakage
150 uA max. @, 264 VAC, 63 Hz
Output
12V DC 0 - 5A
Fuses F1 and F2
Ceramic anti-surge T HBC fuse, 4A 250V rating,
5x20 mm size, marked T4AH250V
2.6
Mains Cordset Specification
The BloodTrack Kiosk V5 uses a locking type mains cordset and should only be
used with these Haemonetics-approved mains cordset types. This table lists
commonly supplied cordsets. For details on cordsets for use in countries outside of