Troubleshooting
Handpiece vibrates, is very noisy
and/or becomes warm in the front
part.
Use of bent burs/tools.
→
Use a
different bur/tool to test this. The
bur/tool itself can be tested with
the Bur Shaft Tester (art. 4990) to
see if it is bent.
Maximum allowed speed of the
tool
was
exceeded,
which
caused a stronger vibration.
→
Pay attention to the tool (e.g. bur,
capper,
and
instrument)
manufacturers handling details.
Bur/tool can be inserted only with
difficulty or not at all.
Use of bent burs/tools.
→
Use a
different bur/tool to test this. The
bur/tool itself can be tested with
the Bur Shaft Tester (art. 4990) to
see if it is bent.
It does not spray.
Pressure cannot be set up in the
tank, because the lid is not
tightened well.
There are parts of dirt in the
spray system.
→
Clean the spray
system
according
to
the
instructions in the manual.
The filter (art. 3497) in the tank is
blocked.
→
Change the filter.
Unit does not work at all.
Check if the main switch of the
unit is turned on at the side of the
unit.
Foot switch is used constantly or
faulty.
→
Pull out the socket of
the foot switch to see if this is the
cause.
Pull out the mains plug to reset
the device.
After lengthy use unit does not
spray.
Unit is overheated.
→
Slots of the
unit should not be covered or
blocked. Let cool down the unit
for a while.
The unit shows different sound
levels at different sites of use.
The sound level varies according
to the surroundings of the unit. If
you set up the unit directly in front
of a wall, the sound is reflected
stronger, as if it would, while
standing further away from a
wall.
The motor of the handpiece
stutters
(e.g.
1
sec.)
while
switching it on.
Very seldom this error might
occur. It would not damage the
motor at all.
Spray stream stutters.
There is air in the spray system.
After 5 min running, this effect will
disappear.
Acceleration of the spray stream.
Note: To reduce the starting time
of the spray, turn the spray
regulator on Max. After the spray
started, adjust the stream to the
requested amount of water.
Fluid leaks out of the unit.
Please be aware of the maximum
filling level of the tank. Do not
exceed the maximum level when
filling the tank.