Smartdop
®
30EX
Vascular Testing Instructions
6.
Allow the sensor to equilibrate. Then, press the
PRINT button or the probe button to begin the test.
7.
Ask the patient to dorsiflex the foot the specified
number of times, following the foot animation on the
LCD. The exercise should be performed forcefully,
especially when lifting the foot upward.
8.
After dorsiflexing, the patient should relax the foot
and avoid all movement. The test is complete
when the waveform on the LCD returns to the
baseline amplitude.
The Smartdop
®
30EX automatically freezes the
venous reflux study waveform. Press the PRINT
button to print the results.
9.
To save the data in memory,
see 5-3-8 “MEMORY -
STORE” for details.
10.
Press the INF/DUMP or probe button (probe button
function must be set to FREEZE or FRZ&PRINT) to
exit FREEZE mode.
11.
To rule out superficial venous reflux and direct
augmentation to the deep veins, the test can be
repeated by placing a vascular cuff above the knee.
Interpreting the Results
A venous refill time of > 20 seconds is considered normal.
Refill time between 18 and 20 seconds is questionable. Repeat the test.
If refill time is less than 18 seconds, a venous condition may exist.
62
5. VENOUS REFLUX (Cont.)
NOTE
Recovery time can range from 1 to 74 seconds
when a count of 5 dorsiflexions is performed for a
maximum test time of 80 seconds.
Eberhardt, RT, Raffeto, JD. Chronic Venous Inssufficiency. Circulation 2005, 111:2398-2409. Retreived November 1, 2013 from
http://circ.ahajournals.org/content/111/18/2398.
WARNING
If any abnormality is found on the unit or patient,
discontinue use in a manner safe to the patient.