4. SETUP & BASIC OPERATION
4-5
MEASURING BLOOD VELOCITY
This section explains the fundamentals of using the
Smartdop
®
30EX to measure blood velocity. For detailed
instructions on performing specific tests, see
Section 2
Testing Instructions
.
1.
Connect the probe to the Smartdop
®
30EX with
the round probe connection mark on the probe
connector pointed up (12 o’clock position).
2.
Press the power button to turn the unit on. Turn the
volume control to the desired level.
3.
Press the MODE button to select Mode 1 for
arterial or Mode 2 for venous.
4.
Place ultrasonic gel on the probe tip or the patients
skin. Make sure there is enough gel to cover the tip
of the probe.
5.
Place the probe on the measurement area at an
angle of 45° to 60° against flow and adjust to locate
the best sounds.
12
Probe connection mark
Power switch
NOTE
Make sure the lighted LED on the probe selector
button corresponds to the probe connector being
used. Press the probe selector button to change
the probe selection.
CAUTION
Use ultrasound gel. Other materials such as baby
oil or cream may produce incorrect sounds or
damage the probe.
WARNING
The ultrasound gel included with purchase is non-
sterile and should only be used noninvasively.
WARNING
Discontinue use of ultrasound gel if an allergic
reaction occurs.
CAUTION
The probe transducer tip is thin and delicate.
Please handle with care and use the probe cap
when not in use.