Turbidity measurement
For accurate turbidity readings use clean sample cells and remove air bubbles.
Measurement notes
Proper measurement techniques are important in minimizing the effects of instrument variation, stray
light and air bubbles. For accurate and repeatable measurements:
Instrument
• Make sure that the instrument is on a level, stationary surface that is free of vibration during the
measurement.
• The USEPA filter assembly is required for turbidity measurements reported for United States
Environmental Protection Agency (USEPA), National Primary Drinking Water Regulations
(NPDWR) or National Pollutant Discharge Elimination System (NPDES) permits.
• Turn the instrument on 30 minutes (Ratio on) or 60 minutes (Ratio off) before measurement. Keep
the instrument on 24 hours a day if the instrument is used regularly.
• Always close the sample compartment lid during measurement, calibration and verification.
• Remove the sample cell from the instrument and turn off the instrument if the instrument is stored
for an extended time period (more than a month).
• Keep the sample compartment lid closed to keep dust and dirt out.
Sample cells
• Always cap the sample cell to prevent spillage of the sample into the instrument.
• Always use clean sample cells in good condition. Dirty, scratched or damaged cells can result in
readings that are not accurate.
• Make sure that cold samples do not “fog” the sample cell.
• Store sample cells filled with distilled or deionized water and cap tightly.
• For the best accuracy, use a single sample cell for every measurement or a flow cell.
Note: As an alternative, matched sample cells may be used for measurements but do not provide as good of
accuracy or precision as a single indexed sample cell or flow cell. When using matched sample cells, align the
orientation mark on the sample cell with the reference mark on the sample cell holder.
Measurement
• Measure samples immediately to prevent temperature changes and settling. Before a
measurement is taken, always make sure that the sample is homogeneous throughout.
• Avoid sample dilution when possible.
• Avoid instrument operation in direct sunlight.
English
17
Содержание TL2300
Страница 2: ...English 3 Fran ais 24 Espa ol 47 Portugu s 69 92 110 131 152 2...
Страница 95: ...A TL2300 90 USB USB 1 1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 95...
Страница 97: ...3 3 1 4 7 2 8 1 3 9 4 10 5 11 6 12 1 4 97...
Страница 98: ...1 105 98 1 2 USB 98...
Страница 102: ...0 1 NTU StablCal StablCal 1 ID 3 2 3 4 USEPA 106 5 USEPA 6 101 7 StablCal 8 9 10 0 1 NTU 11 12 102...
Страница 103: ...13 14 StablCal 7 13 NTU 15 16 Gelex StablCal 1 ID 3 2 3 4 Gelex 5 6 7 8 103...
Страница 104: ...USEPA NPDWR NPDES USEPA 30 60 24 ID ID 100 99 1 ID 3 2 3 ID ID ID 4 30 mL 104...
Страница 105: ...5 6 7 8 9 10 1 2 3 4 ID ID ID ID USB USB 2 0 bmp csv xml 105...
Страница 106: ...1 USB USB 2 3 4 5 105 6 1 2 105 3 1 2 1 2 3 106...
Страница 107: ...30 60 1 2 3 USB USB USB 20 0 RTC 1 3 4 7 8 9 SDRAM 10 NOR 11 SPI 12 13 StablCal 100 101 107 107...
Страница 109: ...99 USB dd mm yyyy hh mm dd mm yyyy hh mm 109...
Страница 112: ...EN 55011 CISPR 11 A IECS 003 A A FCC PART 15 A FCC 15 1 2 FCC 15 A 112...
Страница 113: ...1 2 3 4 5 EMC A A A TL2300 90 1 USB USB 1 113...
Страница 114: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 114...
Страница 115: ...2 1 6 Gelex 2 7 StablCal 3 USEPA 8 AC 4 TL2300 9 5 1 30 mL 6x 10 3100 m 3100 m USB USB 1 114 USB 3 m ID 115...
Страница 116: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 Home Instrument information 1 ID ID Login ID ID ID 4 1 116...
Страница 117: ...1 126 118 117...
Страница 124: ...USEPA 1 NPDWR NPDES USEPA 30 60 1 24 1 124...
Страница 127: ...1 2 1 2 3 30 60 127...
Страница 130: ...Auto Send 118 ID USB ID dd mm yyyy hh mm ID Web dd mm yyyy hh mm ID ID ID 130...
Страница 134: ...EMC A A A TL2300 90 USB USB 1 134...
Страница 135: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 135...
Страница 136: ...2 1 6 Gelex 2 7 StablCal 3 USEPA 8 4 TL2300 9 5 1 30 mL 6 10 3100 m 10 710 ft 3100 m USB USB 1 135 USB 3 m 9 8 ft ID 136...
Страница 137: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 1 ID ID ID ID ID 4 137...
Страница 138: ...1 146 139 138...
Страница 143: ...StablCal 1 Login ID 3 2 Login 3 4 USEPA 148 5 USEPA 6 141 7 StablCal 8 9 10 0 1 NTU 11 12 1 143...
Страница 146: ...5 6 7 8 9 10 Done 1 2 3 Filter Filter Settings 4 146...
Страница 148: ...1 2 3 30 60 1 2 3 USB USB USB 148...
Страница 149: ...20 0 RTC 1 IC 3 4 7 8 9 SDRAM 10 NOR 11 SPI 12 13 StablCal 141 141 148 20 141 StablCal 141 149...
Страница 151: ...dd mm yyyy hh mm ID ID ID 151...
Страница 155: ...1 2 Class A Part 15 FCC Rules 1 2 3 4 5 Korean certification EMC A A A TL2300 Ratio Ratio 90 USB USB 1 155...
Страница 156: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 156...
Страница 157: ...2 1 6 Gelex 2 7 StablCal 3 USEPA 8 4 TL2300 9 5 1 30 6 10 3100 10 710 3100 USB USB 1 156 USB 3 9 8 ID 157...
Страница 158: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 1 Login ID Login Logout ID ID Sample ID ID ID 4 158...
Страница 159: ...1 Verification Data Log 167 Setup 160 Diagnostics Timer 159...
Страница 173: ......