Operation
Configure the instrument settings
1.
Push
, then push
Setup
.
2.
Select an option.
Option
Description
Location
Sets the location name of the instrument. The location is sent with measurements to the
USB drive. The location is not saved to the data log.
Date & Time
Sets the date format, the time format and the date and time. Enter the current date and
time.
Date Format
—Sets the date format. Options: dd-mm-yyyy (default), yyyy-mm-dd,
dd-mm-yyyy or mm-dd-yyyy.
Time Format
—Sets the time format. Options: 12 or
24 hours (default).
Security
Enables or disables password protection for the settings and tasks in the security list.
Security Password
—Sets or changes the security (administrator) password
(10 characters maximum). Passwords are case sensitive.
Security List
—Sets the
security level for each setting and task in the security list.
•
Off
—All operators can change the setting or do the task.
•
One key
—Only operators with a one-key or two-key security level can change the
setting or do the task. Refer to
•
Two keys
—Only operators with a two-key security level can change the setting or do
the task.
Note: The Security setting is not set to on until Close is pushed.
Sound Settings
Enables or disables the sound settings for individual events. To enable or disable all of
the sound settings, select All and then push
Setup
.
Peripherals
Shows the connection status of attached devices such as a printer, USB memory (flash
drive) or keyboard.
Power
Management
Sets when the instrument is automatically set to sleep mode or off after a period of no
activity.
Sleep Timer
—Sets when the instrument is set to sleep mode. Options: OFF,
30 minutes, 1 (default), 2 or 12 hours.
Configure the measurement settings
Select the reading mode, measurement units, data log settings and more.
1.
At the main reading screen, push
Options
>
Reading Setup
.
2.
Select an option.
Option
Description
Reading
Mode
Sets the reading mode to single, continuous or RST mode.
Single (default)
—The
measurement stops when the reading is stable.
Continuous
—The measurement continues
until the user pushes
Done
.
RST
—The Rapidly Settling Turbidity (RST) mode calculates and
continuously updates the turbidity reading of the sample to a confidence of 95%, based on the
accumulated trend of the real time measured values. The RST mode is best used on samples
that settle rapidly and continuously change in value. The reading is based on a correctly
prepared sample that is homogeneous at the beginning of the reading. It is best applied to
samples that are greater than 20 NTU. The sample must be mixed thoroughly by inversion
immediately before inserting it into the instrument.
Signal Avg
—The turbidity reading that
shows on the display is an average of the values measured during the time interval selected.
Options: For single measurement mode, 5 to 15 seconds. For continuous measurement mode,
5 to 90 seconds.
Unit
Selects the measurement units that show on the display and that are recorded to the data log.
Options: NTU (default), EBC.
English
11
Содержание TL2300
Страница 2: ...English 3 Fran ais 24 Espa ol 47 Portugu s 69 92 110 131 152 2...
Страница 95: ...A TL2300 90 USB USB 1 1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 95...
Страница 97: ...3 3 1 4 7 2 8 1 3 9 4 10 5 11 6 12 1 4 97...
Страница 98: ...1 105 98 1 2 USB 98...
Страница 102: ...0 1 NTU StablCal StablCal 1 ID 3 2 3 4 USEPA 106 5 USEPA 6 101 7 StablCal 8 9 10 0 1 NTU 11 12 102...
Страница 103: ...13 14 StablCal 7 13 NTU 15 16 Gelex StablCal 1 ID 3 2 3 4 Gelex 5 6 7 8 103...
Страница 104: ...USEPA NPDWR NPDES USEPA 30 60 24 ID ID 100 99 1 ID 3 2 3 ID ID ID 4 30 mL 104...
Страница 105: ...5 6 7 8 9 10 1 2 3 4 ID ID ID ID USB USB 2 0 bmp csv xml 105...
Страница 106: ...1 USB USB 2 3 4 5 105 6 1 2 105 3 1 2 1 2 3 106...
Страница 107: ...30 60 1 2 3 USB USB USB 20 0 RTC 1 3 4 7 8 9 SDRAM 10 NOR 11 SPI 12 13 StablCal 100 101 107 107...
Страница 109: ...99 USB dd mm yyyy hh mm dd mm yyyy hh mm 109...
Страница 112: ...EN 55011 CISPR 11 A IECS 003 A A FCC PART 15 A FCC 15 1 2 FCC 15 A 112...
Страница 113: ...1 2 3 4 5 EMC A A A TL2300 90 1 USB USB 1 113...
Страница 114: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 114...
Страница 115: ...2 1 6 Gelex 2 7 StablCal 3 USEPA 8 AC 4 TL2300 9 5 1 30 mL 6x 10 3100 m 3100 m USB USB 1 114 USB 3 m ID 115...
Страница 116: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 Home Instrument information 1 ID ID Login ID ID ID 4 1 116...
Страница 117: ...1 126 118 117...
Страница 124: ...USEPA 1 NPDWR NPDES USEPA 30 60 1 24 1 124...
Страница 127: ...1 2 1 2 3 30 60 127...
Страница 130: ...Auto Send 118 ID USB ID dd mm yyyy hh mm ID Web dd mm yyyy hh mm ID ID ID 130...
Страница 134: ...EMC A A A TL2300 90 USB USB 1 134...
Страница 135: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 135...
Страница 136: ...2 1 6 Gelex 2 7 StablCal 3 USEPA 8 4 TL2300 9 5 1 30 mL 6 10 3100 m 10 710 ft 3100 m USB USB 1 135 USB 3 m 9 8 ft ID 136...
Страница 137: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 1 ID ID ID ID ID 4 137...
Страница 138: ...1 146 139 138...
Страница 143: ...StablCal 1 Login ID 3 2 Login 3 4 USEPA 148 5 USEPA 6 141 7 StablCal 8 9 10 0 1 NTU 11 12 1 143...
Страница 146: ...5 6 7 8 9 10 Done 1 2 3 Filter Filter Settings 4 146...
Страница 148: ...1 2 3 30 60 1 2 3 USB USB USB 148...
Страница 149: ...20 0 RTC 1 IC 3 4 7 8 9 SDRAM 10 NOR 11 SPI 12 13 StablCal 141 141 148 20 141 StablCal 141 149...
Страница 151: ...dd mm yyyy hh mm ID ID ID 151...
Страница 155: ...1 2 Class A Part 15 FCC Rules 1 2 3 4 5 Korean certification EMC A A A TL2300 Ratio Ratio 90 USB USB 1 155...
Страница 156: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 156...
Страница 157: ...2 1 6 Gelex 2 7 StablCal 3 USEPA 8 4 TL2300 9 5 1 30 6 10 3100 10 710 3100 USB USB 1 156 USB 3 9 8 ID 157...
Страница 158: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 1 Login ID Login Logout ID ID Sample ID ID ID 4 158...
Страница 159: ...1 Verification Data Log 167 Setup 160 Diagnostics Timer 159...
Страница 173: ......