Make sure to fully rinse the syringe with
methanol dry
before the test to remove titrant. Make sure to
fully rinse the syringe with
methanol dry
after the test to remove water.
Note: A license key is necessary to enable the syringe verification option on the instrument. Contact the
manufacturer or a sales representative to get a license key.
1.
From the home screen, select Settings, then push
Select
.
2.
From the settings screen, select Options, then push
Select
.
3.
Push the left or right arrow key to select the expert mode, then push
Done
.
4.
From the home screen, select Maintenance, then push
Select
.
5.
From the maintenance screen, select Syringe verification, then push
Select
.
6.
When the option is selected for the first time, enter the license key.
7.
Follow the procedure supplied with the syringe verification kit. For additional information, refer to
the expanded version of this manual.
Manage the data log
To select data to view, delete or export, specify data filters
1.
From the home screen, select
Data log
.
2.
Select an option, then push
Select
.
Option
Description
View data log
Views measurement data. Select individual lines of data to view more content.
Export data log
Exports measurement data from the system to an external device. Preview data selection
before it is exported. Make sure that an external device is connected to the instrument
(e.g., a USB key, external hard drive, etc.).
Delete data log
Removes measurement data from the system. Previews data selection before it is
removed.
3.
Specify the data selection parameters. Push the left and right arrow keys to make a selection.
Push the up and down arrow keys to select an option.
Option
Description
Result type
Sets the type of result available.
Application
Sets the available applications.
Date
Sets the date range.
Operator
Sets the available operators.
4.
If
View data log
was the selected option, push
View
to see the selected data.
• Push the up and down arrow keys to select a line of data and push
Detail
to see more data
• If only one
Application
is selected push the up and down arrow keys to select a line of data
and push
Detail
to see more data, or push the left and right arrow keys to see related graphs
5.
If
Export data log
or
Delete data log
was the selected option, push
Preview
to see the selected
data, then push
Export
or
Delete
to start the procedure.
24
English
Содержание TitraLab KF1121
Страница 9: ...English 9...
Страница 14: ...14 English...
Страница 34: ...34 Deutsch...
Страница 39: ...Deutsch 39...
Страница 60: ...60 Italiano...
Страница 65: ...Italiano 65...
Страница 86: ...86 Fran ais...
Страница 91: ...Fran ais 91...
Страница 112: ...112 Espa ol...
Страница 117: ...Espa ol 117...
Страница 138: ...138 Portugu s...
Страница 143: ...Portugu s 143...
Страница 169: ...e tina 169...
Страница 189: ...Dansk 189...
Страница 194: ...194 Dansk...
Страница 214: ...214 Nederlands...
Страница 219: ...Nederlands 219...
Страница 240: ...240 Polski...
Страница 245: ...Polski 245...
Страница 271: ...Svenska 271...
Страница 290: ...290 Suomi...
Страница 295: ...Suomi 295...
Страница 309: ...ESD IECS 003 309...
Страница 311: ...1 6 11 2 2 7 12 1 3 USB 1 8 13 4 9 5 10 1 KF1121 1 2 1 USB USB 311...
Страница 312: ...USB USB USB USB 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 1 1 3 2 4 312...
Страница 313: ...2 1 KF 6 2 KF PTFE 7 3 8 4 USB 9 3 x GL45 5 10 3x 313...
Страница 314: ...15 6 in GFCI GFI 1 2 312 3 180 6 314...
Страница 315: ...315...
Страница 316: ...316...
Страница 317: ...317...
Страница 318: ...3 318...
Страница 319: ...3 1 2 3 MSDS SDS 333 1 2 319...
Страница 320: ...320...
Страница 321: ...4 1 3 2 321...
Страница 322: ...1 3 5 2 4 2 Settings Options 322...
Страница 323: ...1 5 2 6 2 3 KF 7 4 8 2 2 USB 1 2 331 323...
Страница 324: ...MSDS SDS 1 Settings 2 Select 324 3 Back 324...
Страница 326: ...MSDS SDS a 1 2 3 USB 4 332 5 Calibration 6 Select 7 8 9 Start 10 11 326...
Страница 327: ...12 Replicate Sample Skip 13 14 15 16 13 17 Start Injection Autodetect Yes 18 19 Replicate Sample Skip 20 16 Select 327...
Страница 329: ...8 Sample QC Define blank Define blank QC with blank Sample with blank 9 10 Start 11 Cell standby 324 15 12 13 329...
Страница 333: ...300 C 572 F 4 333...
Страница 341: ...Magyar 341...
Страница 346: ...346 Magyar...
Страница 367: ...Rom n 367...
Страница 372: ...372 Rom n...
Страница 392: ...392 T rk e...
Страница 397: ...T rk e 397...
Страница 417: ...Slovensk jazyk 417...
Страница 422: ...422 Slovensk jazyk...
Страница 441: ...Slovenski 441...
Страница 446: ...446 Slovenski...
Страница 465: ...Hrvatski 465...
Страница 470: ...470 Hrvatski...
Страница 484: ...IECS 003 A IECS 484...
Страница 486: ...1 6 11 2 2 7 12 1 3 USB1 8 13 4 9 5 10 1 KF1121 1 2 USB USB USB 1 USB USB 486...
Страница 487: ...USB 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 1 1 3 2 4 487...
Страница 488: ...2 1 KF 6 2 KF PTFE 7 3 8 4 USB 9 3 x GL45 5 10 3x 488...
Страница 489: ...15 cm AC GFCI GFI 1 2 486 3 180 6 489...
Страница 490: ...490...
Страница 491: ...491...
Страница 492: ...492...
Страница 493: ...3 493...
Страница 494: ...3 1 2 3 MSDS SDS o 508 1 2 494...
Страница 495: ...495...
Страница 496: ...4 1 3 2 496...
Страница 497: ...1 ON OFF 3 5 2 4 ON OFF 2 Settings Options 497...
Страница 498: ...1 5 2 6 2 3 KF 7 4 8 2 2 USB 1 2 506 498...
Страница 500: ...Cell management Start USB 1 Home 2 USB USB USB 3 4 Import 5 Home USB 1 Purge 2 All elements Select 3 4 2 5 6 Exit 500...
Страница 501: ...MSDS SDS 1 2 3 USB 4 Cell management 507 5 Calibration 6 Select 7 Information Operator ID 8 9 Start 501...
Страница 508: ...Empty cell pump 2 2 2 Stop pumps Fill cell Empty cell Fill cell Empty cell 3 Exit 300 C 4 508...
Страница 509: ......