Table 10 USB storage device (continued)
Error/Warning
Description
Possible solution
Application {application name}
(untrusted)
This selection is not allowed
because the application has not
been authenticated
Contact technical support to get an authenticated
application file
Application {application name}
(obsolete)
The selection is allowed. When the
application is exported, it will
automatically update to the latest
version
The automatic update will set the new application
parameters to the default values. It is not
recommended to use an obsolete application, but
to install the latest version
Application {application name}
(syringe to replace)
The selection is allowed but the
syringe recommended in the
application file is not the one
installed on the instrument
Change the installed syringe to the one
recommended by the application (refer to
on page 42)
Error while serializing data
Error when creating a file on the
USB storage device
Change or format the USB storage device and try
the export again
Cannot write to USB mass
storage
USB storage device is not installed
Remove the USB storage device and make sure
the USB icon is no longer seen in the header bar,
then install the device again try the export again
USB storage device is corrupt
Change or format the USB storage device and try
the export again
Table 11 Data updates
Error/Warning
Description
Possible solution
Edit user failed. Be sure that name {user
name} is not already used
The update to the operator name
cannot be done
Check that the name is not already used
Fail to save the modification of the
application
The update to the application data
cannot be done
Try again, if the problem continues,
contact technical support
Impossible to remove application
{application name}
The application cannot be erased
Try again, if the problem continues,
contact technical support
Error while deleting data
The data cannot be erased
Try again, if the problem continues,
contact technical support
Table 12 Battery
Error/Warning
Description
Possible solution
Battery voltage low
Battery voltage is below 2.2 V. The date
and time are no longer available
Contact technical support
Realtime clock has been lost The date and time are no longer available
Set the date and time again. If the date and time
are not available the next time the instrument is
set to on, contact technical support
Table 13 Balance
Error/Warning
Description
Possible solution
Format of message received from
the balance is not supported
The data received from the balance is not
compatible with the format of messages
accepted by the instrument
Configure the balance to be compatible
with the format of messages accepted by
the instrument (refer to
on page 19)
No connection with balance
The instrument is not receiving data from
the balance
Check the balance cable is installed
correctly
The measurement received is out
of range
The weight sent by the balance is not in the
acceptance limits
Adjust the sample. Adjust the acceptance
limit for the application
Troubleshooting
50
Содержание TitraLab KF1000 Series
Страница 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Страница 2: ......
Страница 6: ...Specifications 4...
Страница 12: ...General information 10...
Страница 20: ...Installation 18...
Страница 30: ...Startup 28...
Страница 40: ...Standard operations 38...
Страница 54: ...Troubleshooting 52...
Страница 60: ...Replacement parts and accessories 58...
Страница 64: ...Change application parameters 62...
Страница 66: ...Frequently asked questions 64...
Страница 67: ......