Table 10 USB storage device
Error/Warning
Description
Possible solution
Exporting {application name} -
access forbidden to the
applications folder on USB flash
drive
Error when exporting data to the
USB storage device
Make sure that the USB storage device is correctly
installed and the USB icon can be seen in the
header bar (refer to
on page 22)
Remove the USB storage device and make sure
the USB icon is no longer seen in the header bar,
then install the device again and make sure that
the icon can be seen (refer to
on page 22)
Exporting {application name} -
read only access to the
applications folder on USB flash
drive
The USB storage device is read only Unlock the storage device or use another one
Exporting {application name} -
error during writing. Verify USB
flash drive is not full
Error when exporting data to the
USB storage device
Make sure that the USB storage device has
enough available space or use another one
USB flash drive is not installed.
Install and export applications
again
The instrument cannot find a USB
storage device
Make sure that the USB storage device is correctly
installed and the USB icon can be seen in the
header bar (refer to
on page 22)
Remove the USB storage device and make sure
the USB icon is no longer seen in the header bar,
then install the device again and make sure that
the icon can be seen (refer to
on page 22)
Application {application name}:
maximum number of applications
is reached
The maximum number of
applications for each line has been
reached when copying applications
to the instrument from the USB
storage device
Remove any unwanted applications, then push the
Home
key to return to the home screen. Remove
the USB storage device and make sure the USB
icon is no longer seen in the header bar, then
install the device again and import the new
application again
Application {application name}:
requires at least one titrant that
cannot be installed
The titrant needed by the application
cannot be installed on the instrument
because the applications currently
installed use another titrant
Remove all installed applications, then push the
Home
key to return to the home screen. Remove
the USB storage device and make sure the USB
icon is no longer seen in the header bar, then
install the device again and import the new
application again
Application {application name}:
requires too many titrants for this
instrument model configuration
The instrument has a single syringe
and the application needs a two
syringe instrument
Change to a two syringe instrument
Error while saving application
{application name}
Error when copying from the USB
storage device to the instrument
memory
Contact technical support
Unable to read applications from
device. Applications have been
removed, please load them from
USB flash drive
While updating the application tiles
on the display, an installed
application has been detected as
corrupt
Remove the corrupt application and try to load it
again
Application {application name}
corrupted
The selected application file is
invalid
Try to load the original application file again, or
contact technical support for a replacement file
Application {application name}
incompatible with hardware
The hardware does not support the
application
Change to a compatible instrument
Application {application name}
(too new model)
This selection is not allowed
because the instrument software
only supports older application files
Update the instrument software
Troubleshooting
49
Содержание TitraLab KF1000 Series
Страница 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Страница 2: ......
Страница 6: ...Specifications 4...
Страница 12: ...General information 10...
Страница 20: ...Installation 18...
Страница 30: ...Startup 28...
Страница 40: ...Standard operations 38...
Страница 54: ...Troubleshooting 52...
Страница 60: ...Replacement parts and accessories 58...
Страница 64: ...Change application parameters 62...
Страница 66: ...Frequently asked questions 64...
Страница 67: ......