Alarm Relay port
—Connect to external devices, such as horns or lights. Use an alarm relay cable
assembly (6-pin connector on one end and tinned wire leads on the other end) to make the
connection. Refer to
.
Table 10 Modem port wiring
Pin
Wire color
Description
Pin
Wire color
Description
A
Red
Tip
C
—
12 VDC
B
Green
Ring
D
—
12 VDC reference
Figure 6 RJ11-style modular connector adaptor with cover removed
1
Modem cable assembly (2862)
3
Red wire
2
Green wire
4
RJ11 style adaptor (3188)
Table 11 4–20 mA port wiring
Pin
Description
Wire color
Pin
Description
Wire color
A
Output A +
Yellow
C
Output B +
Red
B
Output A –
Black
D
Output B –
Green
Table 12 Alarm Relay 1 and 2 wiring
Pin
Description
Wire color
Pin
Description
Wire color
A
Relay #1 normally open
Green
D
Relay #2 normally open
Green
B
Relay #1 common
Black
E
Relay #2 common
Black
C
Relay #1 normally closed
White
F
Relay #2 normally closed
White
Table 13 Alarm Relay 3 and 4 wiring
Pin
Description
Wire color
Pin
Description
Wire color
A
Relay #3 normally open
Green
D
Relay #4 normally open
Green
B
Relay #3 common
Black
E
Relay #4 common
Black
C
Relay #3 normally closed
White
F
Relay #4 normally closed
White