SECTION 8
TROUBLESHOOTING
8.1
Error Codes
Error codes inform the user of an out-of-range value or meter
problem. Table 6 outlines the operator assistance codes available
in the meter series.
Note: To display the electric current coming from the dissolved oxygen
electrode, press the
READ
and
CONC %
keys simultaneously.
8.2
Meter Service Request Questionnaire
1. What is the complete lot code of the meter and electrode?
2. On what date was the meter purchased?
3. How long has the meter been in use?
4. What types of samples are being tested?
5. What is the temperature of the samples being tested?
6. How often is the meter being used?
7. How is the meter being stored between uses?
8. If the meter has been in use for a while, what maintenance
has been performed?
9. Describe the suspected problem or failure of the meter.
10. Please have your meter, electrode, buffers/standards, and this
completed questionnaire near the phone before calling
technical support.
Table 6
Error Code
Error Type
Possible Remedy
E-1
Data error in the non-volatile memory.
Turn off the meter, then turn it on
again.
E-3
Failure to correctly store a reading.
Call Service. Meter cannot store data
in at least one location, but is
otherwise functional.
E-9
Failure to correctly retrieve a reading that
was stored earlier.
Call Service.
E-10
Sample temperature is out of range
(0 to 50 °C).
NA
Содержание sension8
Страница 6: ...6 ...
Страница 8: ...8 ...
Страница 10: ...10 ...
Страница 12: ...12 ...
Страница 20: ...20 ...
Страница 44: ...44 ...
Страница 48: ...48 ...
Страница 50: ...50 ...
Страница 64: ...64 ...
Страница 76: ...76 ...
Страница 78: ...78 HOW TO ORDER ...
Страница 79: ...79 REPAIR SERVICE ...
Страница 84: ......