29
SECTION 3,
continued
3.1.10 Automatic Data Transfer
This setup feature activates the meter’s automatic data transfer
(Print) function.
The automatic data transfer function automatically sends data
through the RS232C port, depending upon the time interval
selected. Time intervals are selected from the following options:
10 seconds, 30 seconds, 1 minute, 5 minutes, 20 minutes, 1 hour,
2 hours, or 6 hours.
Accessing the calibration mode or the setup mode halts automatic
data transfer. Also, if the meter has been set to the Lock mode
using setup 1, the meter will not send data. When the meter is in
Lock mode and the
READ
key is pressed, automatic data transfer
will occur at selected time intervals only until the meter stabilizes
and the value in the display is locked.
To change the automatic data transfer setup:
1. Press the
SETUP
key.
2. Press the
UP ARROW
nine times. A
10
will begin flashing in
the upper right hand corner of the display. The current
automatic data transfer entry will be displayed.
3. Press the
EXIT
key to leave the value unchanged. To change
the value, press the
ENTER
key. The flashing question mark
will appear in the upper right corner of the display next to
the 10. Use the arrow keys to view the time intervals for
automatic data transfer.
4. Press the
ENTER
key and then the
EXIT
key when the desired
automatic data transfer time interval appears in the display.
The meter will return to the Read Mode.
Each time data transfer occurs, the Print icon will momentarily
appear at the top of the display.
Содержание sension8
Страница 6: ...6 ...
Страница 8: ...8 ...
Страница 10: ...10 ...
Страница 12: ...12 ...
Страница 20: ...20 ...
Страница 44: ...44 ...
Страница 48: ...48 ...
Страница 50: ...50 ...
Страница 64: ...64 ...
Страница 76: ...76 ...
Страница 78: ...78 HOW TO ORDER ...
Страница 79: ...79 REPAIR SERVICE ...
Страница 84: ......