65
SECTION 7
TROUBLESHOOTING
7.1
Error Codes
Error codes inform the user of an out-of-range value or meter
problem.
outlines the operator assistance codes available
in the meter series.
Note:
To display the electric current coming from the dissolved oxygen
electrode, press the
READ
and
CONC %
keys simultaneously.
Table 6
Error Code
Error Type
Possible Remedy
E-1
Data error in the non-volatile memory.
Turn off the meter, then turn it on
again.
E-3
Failure to correctly store a reading.
Call Service. Meter cannot store data
in at least one location, but is
otherwise functional.
E-9
Failure to correctly retrieve a reading that
was stored earlier.
Call Service.
E-10
Sample temperature is out of range
(0 to 50 °C).