SECTION 3,
continued
4.
Press the
READ ENTER
key three times to skip to the display
showing
100%
.
5.
Press the
READ ENTER
key. The stabilizing icon will appear
while the meter completes the calibration.
6.
When the calibration is complete, the meter will return to the
read mode. Press the
EXIT
key during the calibration sequence
to back out of the calibration routine, one screen at a time,
without completing a calibration.
3.3.7
Calibration Review
To review the last calibration:
1.
Press the
REVIEW
key on the keypad. The date and year of the
last calibration will show.
2.
Press the
TIME
key on the keypad to view the time of the
last calibration.
3.
Press the up arrow. The dissolved oxygen concentration of
calibration will show.
4.
Press the
CONC %
key to view the % saturation and mg/L
values of calibration.
5.
Press the up arrow key. The barometric pressure entry of
calibration will show.
6.
Press the up arrow key. The altitude entry of calibration
will show.
7.
Press the up arrow key. The salinity entry of calibration will
show. Press the
EXIT
key to leave the calibration review.