19
SYSTEM
SYSTEM
Sample ID
Data output
Type Of Report
Sample ID
Data output
Type Of Report
SYSTEM
Data output:
Deactivated
For printer
For computer
External net
SYSTEM
Data output:
Deactivated
For printer
For computer
External net
SYSTEM
PC data output:
Terminal
ComLabo
ComLabo Easy
10-04-2010 17:45
SYSTEM
°C °F
KCl
1D
(mS/cm)
KCl 0.1D
(mS/cm)
KCl 0.01D
(µS/cm)
NaCl 0.05%
(µS/cm)
0
32
65.14
7.13
773
540.40
1
33.8
66.85
7.34
796
557.73
2
35.6
68.58
7.56
820
575.20
3
37.4
70.32
7.77
843
592.79
4
39.2
72.07
7.98
867
610.53
5
41
73.84
8.20
891
628.40
6
42.8
75.62
8.42
915
646.40
7
44.6
77.41
8.64
940
664.55
8
46.4
79.21
8.86
965
682.83
9
48.2
81.03
9.08
989
701.26
10
50
82.85
9.31
1014
719.82
11
51.8
84.68
9.54
1039
738.53
12
53.6
86.54
9.76
1065
757.37
13
55.4
88.39
9.99
1090
776.36
14
57.2
90.26
10.22
1116
795.48
15
59
92.13
10.46
1142
814.74
16
60.8
94.02
10.69
1168
834.14
17
62.6
95.91
10.93
1194
853.68
18
64.4
97.81
11.16
1220
873.36
19
66.2
99.72
11.40
1247
893.18
20
68
101.63
11.64
1273
913.13
21
69.8
103.56
11.88
1300
933.22
22
71.6
105.49
12.12
1327
953.44
23
73.4
107.42
12.36
1354
973.80
24
75.2
109.36
12.61
1381
994.28
25
77
111.31
12.85
1409
1014.90
26
78.8
113.27
13.10
1436
1035.65
27
80.6
115.22
13.35
1464
1056.53
28
82.4
13.59
1491
1077.54
29
84.2
13.84
1519
1098.67
30
86
14.09
1547
1119.92
31
87.8
14.34
1575
1141.30
32
89.6
14.59
1603
1162.80
33
91.4
14.85
1632
1184.41
34
93.2
15.10
1660
1206.15
35
95
15.35
1688
1228.00
36
96.8
15.61
1717
1249.96
37
98.6
15.86
1745
1272.03
38
100.4
16.12
1774
1294.96
39
102.2
16.37
1803
1316.49
40
104
16.63
1832
1338.89
41
105.8
16.89
1861
1361.38
42
107.6
17.15
1890
1383.97
43
109.4
17.40
1919
1406.66
44
111.2
17.66
1948
1429.44
45
113
17.92
1977
1452.32
46
114.8
18.18
2007
1475.29
47
116.6
18.44
2036
1498.34
48
118.4
18.70
2065
1521.48
49
120.2
18.96
2095
1455.71
50
122
19.22
2124
1568.01
ENGLISH
4. Operation
4.14.2. Demal (D) and NaCl 0.05% standards:
Table of conductivity values according to temperature
Values stored in the memory of
sensION
™
+
EC 71
.
5. Sending data via the RS 232 C
Specifications
Baud Rate: 9600 bps
Word length: 8 bits
Stop bits: 2 bits
Parity: None
How to activate communication
At the start-up of the instrument or from “System”, see page 17.
Data sending to a printer
(thermic or dot printer)
40 columns printers.
Thermic: Compatible EPSON (ESC/POS emulation)
Dot-Matrix: Compatible CITIZEN
Data sending to PC, Hyperterminal.
Data sending through the RS232C in printer format.
With the standard application Hyperteminal from Windows, the
data can be captured in the same format as the printer reports,
see pages 11 and 12. They can be exported to an Excel program,
as well.
See PC connection cables at page 5.
EXTERNAL net
Specific communication used when the
sensION
™
+
EC 71
is
connected to a external titration system.
Data sending to PC, ComLabo software
ComLabo (PN LZW8999.99). ComLabo
Software allows control-
ling from a PC several modules, pH-meters, conductivity meters,
automatic burettes, Sampler, etc.
ComLabo Easy
(PN LZW8997.99). ComLabo Easy
Software allows
obtaining data of pH-meters and conductivity meters from a PC.
Main features shared
Windows based software.
Data Logger. All calibration and measuring data are stored in
the PC.
Sample code assignment from PC
Data export to EXCEL.