15.
On the
Controller organizer
tab, select
Tasks
>
Main Task
>
MainProgram
>
MainRoutine
.
For each parameter of the simulation, add a COP module with the data that follow (refer to the
table below):
• Source—Add the controller name, then the I.Data based on the bytes size.
• Dest—Add the parameter Tag.
• Length—Add the bit size of the Tag.
Tags
Type
Size
I.Data
1 - Simulation_value
Float (Real)
4 bytes
0
2 - Simulation_Temperature
Float
4 bytes
4
3 - Simulation_Heartbeat
Integer (INT)
2 bytes
8
4 - Simulation_Error
Integer
2 bytes
10
5 - Simulation_Status
Integer
2 bytes
12