![Hach POLYMETRON 8810 ISE Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/hach/polymetron-8810-ise/polymetron-8810-ise_basic-user-manual_577099018.webp)
The 8810 configuration is:
• Speed: 9600 baud
• Data: 8 bits
• Stop bit: 2
• Parity: none
On startup the analyzer sends the name and software version:
•
ANALYZER 8810 : 00 : 00 : 00 :
•
pH X.XX : 00 : 00 : 00 :
In measurement mode, the analyzer sends the following data:
•
Titration HH : MM : SS
•
M1 XXXXEXX XX.X°C XXX
for the first final point
•
M2 XXXXEXX XX.X°C XXX
for the second final point
•
M3 XXXXEXX XX.X°C XXX
for the differential between the two final points
Where:
• MX XXXXEXX = concentration measurement value
• XX.X°C = temperature
• XXX = value of the potential
In graph plotting mode, the analyzer sends the following data:
•
Plotting HH : MM : SS
•
P : XX.X°C XXX
Where:
• XX.X°C = temperature
• XXX = pH value
Where appropriate, the analyzer also sends the following messages with a time stamp (HH : MM :
SS):
ACTIONS
Cleaning
Chemical cleaning
Stopped
Analyzer stopped
Fix-time
Time between two cycles
Standby
Analyzer in standby
AUTO SYS CAL
Automatic calibration
PROCESS SYS CAL
Process calibration
MANUAL SYS CAL
Manual calibration
SYSTEM ERRORS
LEVEL SAMPLE
Sample missing from measuring cell
OVER TITR TIME
Titration time is too long
ERROR CALIB
Automatic calibration error