background image

Section 1 Product overview

The Intellical PHC301 series pH probes are digital, combination electrodes that measure the pH of
wastewater, drinking water and general water samples. The probes are refillable and have a built-in
temperature sensor and a double ceramic pin for a reference junction. A 28-mL bottle of reference
electrolyte filling solution is supplied with the probe. Refer to 

Figure 1

.

Note: Do not use the probe for measurements in organic solvents.

Figure 1  Probe overview

1   

Probe soaker bottle with storage solution

5   

Cable

2   

Glass bulb, reference junction and
temperature sensor

6   

Probe soaker bottle holder

3   

Filling-hole plug

7   

Electrode filling and storage solution

4   

Protective tape and filling hole

Section 2 Specifications

Specifications are subject to change without notice.

Specifications

Details

Probe type

Digital combination pH probe with refillable electrolyte and built-in
temperature sensor

pH range

0 to 14 pH

pH accuracy

±0.02 pH

Reference type

Ag/AgCl

Reference junction

Double ceramic porous pin

Slope

–59 mV/pH (90 to 110% at 25 °C (77 °F) per Nernstian theoretical
value)

Isopotential point

0 (±30) mV at 7.0 (±0.5) pH

Sodium (alkalinity) error

–0.6 pH at pH 12.6 in 1 M NaOH

Temperature accuracy

±0.3 °C (±0.54 °F)

Temperature sensor type

30 kΩ NTC thermistor

Operating temperature

0 to 50 °C (32 to 122 °F)

English

   

3

Содержание PHC301

Страница 1: ...DOC022 53 80032 PHC301 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration notes 7 5 2 Calibration procedure 8 Section 6 Sample measurement 9 6 1 Sample measurement notes 9 6 2 Sample measurement procedure 9 Section 7 Verify the calibration 10 7 1 Verification procedu...

Страница 4: ...Table of Contents 2...

Страница 5: ...ction and temperature sensor 6 Probe soaker bottle holder 3 Filling hole plug 7 Electrode filling and storage solution 4 Protective tape and filling hole Section 2 Specifications Specifications are su...

Страница 6: ...FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellical probes do not tr...

Страница 7: ...e protective tape from the filling hole of new probes A probe with a blocked filling hole will not operate correctly New probes are filled with electrolyte filling solution and have a soaker bottle th...

Страница 8: ...Figure 2 Remove the protective tape 6 English...

Страница 9: ...ore pH units apart For a one point calibration use a pH buffer near the expected sample pH Use the default calibration options or change the options in the probe settings menu Use the single display m...

Страница 10: ...the probe in the first pH buffer solution Make sure that the sensor and reference junction are fully in the solution Do not put the probe on the bottom or sides of the beaker 6 Shake the probe from s...

Страница 11: ...a point and when the meter is set to read at regular intervals The user must manually save each data point when the meter is set to read continuously Air bubbles below the sensor can cause a slow resp...

Страница 12: ...he verification menu The display shows the pH buffer solution to use for verification Note Menu name for HQd meters Run check standard 2 Prepare or pour the pH buffer solution into a beaker 3 Open the...

Страница 13: ...ater Dry the probe body with a lint free cloth 4 Soak the probe in pH 4 buffer for 20 minutes 5 Rinse the probe with deionized water Dry the probe body with a lint free cloth Table 1 Cleaning solution...

Страница 14: ...he probe 8 3 Replace the filling solution If the filling solution becomes contaminated replace the filling solution 1 Tilt the probe and open the filling hole 2 Use a plastic transfer pipet to remove...

Страница 15: ...age solution or 3 M KCl Do not store the probe in deionized water or in samples of low ionic strength For short term storage the probe can stay in the sample for a maximum of 2 hours if the sample pH...

Страница 16: ...ilization or inaccurate measurements The sample absorbs carbon dioxide CO2 from the air which causes the pH value to slowly decrease in low ionic strength LIS or high purity samples Use the LIS chambe...

Страница 17: ...ctions 250 mL C20C380 10 1 Recommended standards Description Unit Item no pH 4 01 buffer solution Singlet one use packets 20 mL each 20 pkg 2770020 pH 7 00 buffer solution Singlet one use packets 20 m...

Страница 18: ...0 mL S11M010 pH buffer 9 23 technical 500 mL S11M011 10 2 Accessories Description Quantity Item no Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plas...

Страница 19: ......

Страница 20: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: