Section 8 Maintenance

Regular maintenance is necessary for the best accuracy, stabilization time and life of the probe.
Keep the probe in the recommended storage solution between measurements.

8.1 Clean the probe

Clean the probe regularly to remove contamination and to keep the reference junction open.
Symptoms of contamination:

• Incorrect or irregular readings
• Slow stabilization times
• Calibration errors
• Sample material stays on the probe

1.

Rinse the probe with deionized water. Use warm (35–45 °C (95–113 °F)) deionized water to
remove storage solution that dries on the probe. Dry the probe body with a lint-free cloth.

2.

Soak the glass bulb and reference junction in the applicable cleaning solution for the specified
time. Refer to 

Table 1

 and 

Consumables

 on page 15.

3.

Rinse or soak the probe for 1 minute in deionized water. Dry the probe body with a lint-free cloth.

4.

Soak the probe in pH 4 buffer for 20 minutes.

5.

Rinse the probe with deionized water. Dry the probe body with a lint-free cloth.

Table 1  Cleaning solution

Contamination

Cleaning solution

Active component

Soak time

General
contamination

Electrode cleaning solution
for regular maintenance

KATHON

 CG,

DECONEX

®

11 

12–16 hours

Minerals

Electrode cleaning solution
for minerals/inorganic
contamination

Phosphoric acid
(~10%)

10–15 minutes

Fats, grease and oils

Electrode cleaning solution
for fats, oils and grease
contamination

KATHON

 CG,

TRITON

®

 X

2 hours maximum

Proteins

Electrode cleaning solution
for proteins/organic
contamination

Pepsin in HCl

3 hours maximum

Wastewater and
organic compounds

Electrode cleaning solution,
extra strong

Sodium hypochlorite

5–10 minutes

Blockage of porous
pin/diaphragm
reference junctions

Electrode cleaning solution
for porous pin/diaphragm
reference junctions

Thiourea

3 hours with stirring
or until the porous
plug becomes white

English

   

11

Содержание PHC301

Страница 1: ...DOC022 53 80032 PHC301 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration notes 7 5 2 Calibration procedure 8 Section 6 Sample measurement 9 6 1 Sample measurement notes 9 6 2 Sample measurement procedure 9 Section 7 Verify the calibration 10 7 1 Verification procedu...

Страница 4: ...Table of Contents 2...

Страница 5: ...ction and temperature sensor 6 Probe soaker bottle holder 3 Filling hole plug 7 Electrode filling and storage solution 4 Protective tape and filling hole Section 2 Specifications Specifications are su...

Страница 6: ...FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellical probes do not tr...

Страница 7: ...e protective tape from the filling hole of new probes A probe with a blocked filling hole will not operate correctly New probes are filled with electrolyte filling solution and have a soaker bottle th...

Страница 8: ...Figure 2 Remove the protective tape 6 English...

Страница 9: ...ore pH units apart For a one point calibration use a pH buffer near the expected sample pH Use the default calibration options or change the options in the probe settings menu Use the single display m...

Страница 10: ...the probe in the first pH buffer solution Make sure that the sensor and reference junction are fully in the solution Do not put the probe on the bottom or sides of the beaker 6 Shake the probe from s...

Страница 11: ...a point and when the meter is set to read at regular intervals The user must manually save each data point when the meter is set to read continuously Air bubbles below the sensor can cause a slow resp...

Страница 12: ...he verification menu The display shows the pH buffer solution to use for verification Note Menu name for HQd meters Run check standard 2 Prepare or pour the pH buffer solution into a beaker 3 Open the...

Страница 13: ...ater Dry the probe body with a lint free cloth 4 Soak the probe in pH 4 buffer for 20 minutes 5 Rinse the probe with deionized water Dry the probe body with a lint free cloth Table 1 Cleaning solution...

Страница 14: ...he probe 8 3 Replace the filling solution If the filling solution becomes contaminated replace the filling solution 1 Tilt the probe and open the filling hole 2 Use a plastic transfer pipet to remove...

Страница 15: ...age solution or 3 M KCl Do not store the probe in deionized water or in samples of low ionic strength For short term storage the probe can stay in the sample for a maximum of 2 hours if the sample pH...

Страница 16: ...ilization or inaccurate measurements The sample absorbs carbon dioxide CO2 from the air which causes the pH value to slowly decrease in low ionic strength LIS or high purity samples Use the LIS chambe...

Страница 17: ...ctions 250 mL C20C380 10 1 Recommended standards Description Unit Item no pH 4 01 buffer solution Singlet one use packets 20 mL each 20 pkg 2770020 pH 7 00 buffer solution Singlet one use packets 20 m...

Страница 18: ...0 mL S11M010 pH buffer 9 23 technical 500 mL S11M011 10 2 Accessories Description Quantity Item no Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plas...

Страница 19: ......

Страница 20: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: