PHC108

05/2021, Edition 3

User Manual

Содержание PHC108

Страница 1: ...DOC022 53 80428 PHC108 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample requirements 7 6 2 Sample measurement notes 7 6 3 Sample measurement procedure 8 Section 7 Verify...

Страница 4: ...Table of Contents 2...

Страница 5: ...e overview 1 Probe soaker bottle with storage solution 3 Reference junction 2 Cable 4 Glass sensor and temperature sensor Section 2 Specifications Specifications are subject to change without notice S...

Страница 6: ...CC ISED Section 3 Safety information 3 1 Intended use The Intellical PHC108 probe is intended for use by individuals who measure the pH of semi solid samples emulsions and slurries in the laboratory T...

Страница 7: ...ith a soaker bottle that contains storage solution to keep the glass sensor and reference junctions hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...on the sensing element Make sure that the probe stays clean if proteins are in the sample Do not measure samples that are hotter or colder than the specified operating temperature of the probe Refer t...

Страница 10: ...ffer solution to make sure the result is accurate The meter compares the selected pH buffer value to the measured pH value and accepts or rejects the measurement The user can change the pH buffer solu...

Страница 11: ...robe Dry the probe body with a lint free cloth 2 Soak the glass bulb and reference junction in the applicable cleaning solution for the specified time Refer to Table 1 and Consumables on page 11 3 Rin...

Страница 12: ...ch solution 2 Rinse the probe with deionized water Blot dry with a lint free cloth 3 Calibrate the probe 8 3 Storage N O T I C E Probes can become permanently damaged if kept in a storage solution tha...

Страница 13: ...les Increase the sample temperature or adjust the temperature of different samples to be the same within 2 C 3 6 F Procedure problem causes slow stabilization and prevents accurate calibrations or mea...

Страница 14: ...solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769920 pH 7 00 and 10 01 buffer solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769820 pH color coded buffer solution kit NIST 5...

Страница 15: ...SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker 30 mL plastic green 80 pkg SM5014 Beaker dispenser and holder 30...

Страница 16: ...14 English...