Run check standard

The run check standard feature validates instrument performance between sample 
measurements. Use the run check standard feature for periodic or user-defined interval 
measurement of a traceable standard solution. Set the criteria for check standards from 
the MTC101 settings menu.

Note: Access control must be off or a valid password entered before any of the check standard 
method options can be changed.

1.

Select 

Run Check Standard

 from the Meter Options menu.

Note: Select the appropriate probe if two probes are connected to the meter.

2.

Obtain the standard solution specified in the display.

3.

Place the electrode in the standard solution and select 

READ

. The display will show 

Stabilizing... 

and a progress bar as the electrode stabilizes in the standard. The 

display shows the value of the check standard and either 

Check Standard Passed 

or 

probe 

Check Standard Failed

.

If the display reads 

Check Standard Passed

, the check standard measurement is 

within the accepted limits defined by the administrative user. Select 

DONE

 to proceed 

with sample measurement.

If the display reads 

Check Standard Failed

, the measurement is outside of accepted 

limits defined by the administrative user and a recalibration is suggested. If the 
acceptance criteria is set to 

Cal Expires on Failure: Yes

, the display will show the 

calibrate icon with a question mark until the electrode is recalibrated. Recalibrate the 
electrode to correct the status indicator (

Calibration on page 3

).

Measurement

Measurement notes

Data is stored automatically in the data log when 

PRESS TO READ

 or 

INTERVAL 

is 

selected in the 

Setup Measurement Mode

To significantly reduce the response time for reducing-type samples:

The platinum disc must be clean and smooth.

Immerse the platinum disc for 3-10 minutes in 

Reducing Solution for ORP 

Electrodes

 prior to sample analysis.

Rinse the electrode with sample, then measure.

Before measurement:

The electrode must have the correct service-life time stamp. 

Set the date and time in the meter before attaching the 

probe.

The probe is factory calibrated and ready for use. For applications that require greater accuracy and precision, verify the 
user calibration with check standards or perform a user calibration. Routine calibrations are required (

Calibration on page 3

).

Make sure that the platinum electrode is clean before sample measurement (

Clean the probe on page 9

).

If complete traceability is required, enter a sample ID and operator ID before measurement.

Refer to the troubleshooting section for measurement errors (

Troubleshooting on page 13

).

Stir samples gently during measurement.

When deploying the rugged probe at a distance, toss the cell body with a gentle underhand throw. Swinging the probe by the 
cable may cause injury to the user, will cause severe strain on the cable and will shorten the service life of the cell. Damage 
under these conditions may not be covered by the product warranty.

Содержание MTC10101

Страница 1: ...EC European electrical equipment users must now return old or end of life equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment pr...

Страница 2: ...int free cloth 4 If the electrode will not be used immediately fill the storage cap halfway with Hach Electrode Storage Solution or 3M KCl and install the cap 5 Install the shroud and locking ring and...

Страница 3: ...be must have the correct service life time stamp Set the date and time in the meter before attaching the probe It is not necessary to recalibrate when moving a probe from one HQd meter to another if t...

Страница 4: ...er blot dry and place it in the standard solution Make sure that the sample level is above the reference junction 5 Push Read Stir gently The display will show Stabilizing and a progress bar as the pr...

Страница 5: ...estion mark until the electrode is recalibrated Recalibrate the electrode to correct the status indicator Calibration on page 3 Measurement Measurement notes Data is stored automatically in the data l...

Страница 6: ...nect the electrode to the meter 3 Rinse the electrode with the sample and place the electrode in fresh sample 4 Push Read The display will show Stabilizing and a progress bar as the probe stabilizes i...

Страница 7: ...ng many substances in solution A redox titration consists of adding to an unknown sample small increments of a titrant that converts the unknown to a different oxidation state After each addition of t...

Страница 8: ...ng the potential after each addition Near the end point when large potential changes appear add increments of 0 1 0 2 mL Continue the titration 3 4 mL past the end point 5 Plot the electrode potential...

Страница 9: ...d 1 Unscrew the locking ring and slide the shroud and locking ring off the probe 2 Slide the shroud and locking ring off the probe To install the shroud 1 Place the locking ring on the probe body with...

Страница 10: ...ings which cannot be changed A saved method can be used instead of repeatedly adjusting individual settings Changes made to a user defined method are automatically saved with the existing name Multipl...

Страница 11: ...V The measurement limits can be set to match the acceptable values for the sample When the measurement is above the upper limit setting or below the lower limit setting the meter will show an Out of l...

Страница 12: ...200 0 mV Custom standard value Custom standards are characterized at 25 C Table 4 Calibration options settings Settings Option Description Table 5 Check standard options settings Settings Options Desc...

Страница 13: ...sponse time Contaminated reference junctions Rinse reference junction with deionized water thoroughly shake the electrode downward to remove any air bubbles and soak in 3 0 M KCl storage solution for...

Страница 14: ...ts Check standard value is outside of limits set in the current method Verify that the standard is within the limits of the current method Create a new method that expands the acceptable limits Measur...

Страница 15: ...storage cap lid 1 5829800 Field kit Includes glove kit and 5 120 mL sample cups 1 5825800 Glove kit only for HQd meters 1 5828700 IntelliCAL probe stand for standard probes 1 8506600 Meter stand 1 850...

Страница 16: ...insert for meter and two probe storage cavities space for storage of HQd meter with protective glove 1 8505600 Rugged shroud kit 1 5825900 Consumables Electrode cleaning solution 500 mL 2965249 Electr...

Отзывы: