![Hach MP-4 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/hach/mp-4/mp-4_user-manual_577073028.webp)
Operation
28
3.
Push the
MR/DOWN
key until Linc
displays.
4.
Push the
CAL/MCLR
key. SEL displays with the
User
icon.
Note:
Any additional display of KCl, NaCl or 442 icons indicates a Linc between the
additional solution and the User solution. If none of the solution selection icons are
displayed, nothing is linked to the User mode.
5.
Push
UP/MS
or
MR/DOWN
to select a standard solution to link to the User
mode calibration constant.
6.
Push the
CAL/MCLR
key to accept the setting. The User mode now uses the
calibration offset constant that was created here.
Note:
To exit without changing the setting, push any measurement key.
3.20.3 Cancel User mode calibration Linc
To cancel the User mode calibration Linc:
Note:
The MP series meter must be in User linked mode to cancel the “Linc.”
1.
Push a (linked) measurement key such as
COND
,
RES
,
MIN/SALT
,
or
TDS
.
Two solutions are displayed on the left side of the display: User and another,
such as KCl.
2.
Push the
MR/DOWN
key until Linc
displays.
3.
Push the
CAL/MCLR
key. SEL, User and the linked solution appear on the
display.
Содержание MP-4
Страница 2: ...2 ...
Страница 6: ...Table of contents 6 ...
Страница 10: ...Specifications 10 ...
Страница 30: ...Operation 30 ...
Страница 36: ...Calibration 36 ...
Страница 42: ...Troubleshooting 42 ...
Страница 48: ...Limited Warranty 48 ...
Страница 52: ...Conductivity conversion 52 ...
Страница 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...