Operation
19
5.
Push the
pH
key.
6.
Observe or record the value displayed, or push
UP/MS
to store the reading.
Important Note:
After the test, fill the pH/ORP sensor cup with pH Storage Solution
and replace the protective cap. Do not allow the pH/ORP sensor cup to dry out.
3.11 Select a solution
Conductivity, resistivity and TDS (including mineral/salt) require temperature
compensation to 25 °C. The solution profile selection determines the temperature
compensation of conductivity and calculation of TDS and mineral/salt from
compensated conductivity.
There are four solution types:
•
KCl
•
NaCl
•
442
•
User
On the left side of the display is the salt solution characteristic used to model
temperature compensation of conductivity and its TDS conversion. By default, KCI
is used for conductivity, NaCl is used for resistivity (and mineral/salt), and 442
(natural water characteristic) is used for TDS. The User selection allows a custom
value to be entered for the temperature compensation of conductivity and the
conversion ratio if measuring TDS.
Check the display to see if the solution profile displayed is the solution type desired
for that measurement. To change a solution:
1.
Push the
COND
key, the
RES
key, the
MIN/SALT
key or the
TDS
key to select
the parameter to change the solution type.
2.
Push and hold the
CAL/MCLR
key for three seconds and wait for
SEL
to
appear in the display.
3.
Push
UP/MS
or
MR/DOWN
to scroll to the desired solution type.
4.
Push the
CAL/MCLR
key to accept the new solution.
3.11.1Temperature compensation
Electrical conductivity indicates solution concentration and ionization of the
dissolved material. Because temperature affects ionization, conductivity
measurements change with temperature and must be corrected to read at 25
º
C.
Temperature compensation uses the characteristics of salt solutions. The selected
salt solution is displayed on the left side of the display. By default, the meter uses
KCl for conductivity, NaCl for resistivity and 442 for TDS (see
The User mode customizes the temperature compensation and the conversion ratio
if measuring TDS.
Note:
Calibration of each solution type is performed separately and calibration of one solution
does not affect the calibration of the other solution types.
Содержание MP-4
Страница 2: ...2 ...
Страница 6: ...Table of contents 6 ...
Страница 10: ...Specifications 10 ...
Страница 30: ...Operation 30 ...
Страница 36: ...Calibration 36 ...
Страница 42: ...Troubleshooting 42 ...
Страница 48: ...Limited Warranty 48 ...
Страница 52: ...Conductivity conversion 52 ...
Страница 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...