Operation
18
2.
Hold the meter at a 30 degree angle and let the sample flow into the
conductivity cup continuously with no aeration.
3.
Push the
RES
key.
4.
Observe or record the value displayed.
Note:
If reading is lower than 10 k
Ω
, [- - - -] is shown. Measure conductivity for these samples.
3.7
Measure mineral/salt (MP-6p model only)
To measure mineral/salt:
1.
Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2.
Fill the conductivity cup with the solution.
3.
Push the
MIN/SALT
key.
4.
Observe or record the value displayed, or push
UP/MS
to store the reading.
3.8
Measure TDS
To measure TDS:
1.
Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2.
Fill the conductivity cup with the solution.
3.
Push the
TDS
key.
4.
Observe or record the value displayed, or push
UP/MS
to store the reading.
3.9
Measure ORP/Redox (MP-6 and MP-6p models)
To measure the ORP/Redox:
1.
Remove the protective pH/ORP sensor cap. Squeeze the sides and pull up.
2.
Rinse the sensor cup three times with the sample to be measured.
3.
Shake the meter after each rinse to remove residual liquid.
4.
Fill both sensor cups with the sample.
5.
Push the
ORP
key.
6.
Observe or record the value displayed, or push
UP/MS
to store the reading.
Important Note:
After the test, fill the pH/ORP sensor cup with pH Storage Solution
and replace the protective cap. Do not allow the pH/ORP sensor cup to dry out.
3.10 Measure pH (MP-6 and MP-6p models)
To measure pH:
1.
Remove the protective pH/ORP sensor cap. Squeeze the sides and pull up.
2.
Rinse the pH/ORP sensor cup three times with the sample to be measured.
3.
Shake the meter after each rinse to remove residual liquid.
4.
Fill both sensor cups with the sample.
Содержание MP-4
Страница 2: ...2 ...
Страница 6: ...Table of contents 6 ...
Страница 10: ...Specifications 10 ...
Страница 30: ...Operation 30 ...
Страница 36: ...Calibration 36 ...
Страница 42: ...Troubleshooting 42 ...
Страница 48: ...Limited Warranty 48 ...
Страница 52: ...Conductivity conversion 52 ...
Страница 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...