61
Modbus register maps
A.14 Last sample data
shows the register block mirrors of the real-time and buffered data register
blocks with different data. This block is updated with the most recent data at the end of
each sample. Data remains available until the next sample. The update interval is based
on the sample and hold time programmed into the configuration registers (
Configuration information on page 54
1463
Key/Passphrase length
R/W
2 bytes
0-48 WPA/WPA2 or
0-63 WEP
1464
Roaming
R/W
2 bytes
Disabled, Enabled
1465-1472
Radio firmware version
R
16 bytes
ASCII String
1473-1499
Expansion
R
Table 29 Wireless configuration (continued)
Address
Register description
Access
Size
Notes
Table 30 Last sample data
Address
Register Description
Access
Size (bytes) Notes
1500
Sample UTC timestamp—year
R
2
YYYY (2000–9999)
1501
Sample UTC timestamp—month/day
R
2
MD (1–12, 1–31)
1502
Sample UTC timestamp—hour
R
2
H (0–23)
1503
Sample UTC timestamp—minute/second
R
2
MS (0–59, 0–59)
1504
Sample period—hours
R
2
H (0–23)
1505
Sample period—minutes/seconds
R
2
MS (0–59, 0–59)
1506
Reserved
1507–1508
Sample volume
R
4
1509
Sample status
R
2
Bitmap
1510
Reserved
1511
Size 1 counts
R
4
0–4,294,967,295
1513
Size 2 counts
R
4
0–4,294,967,295
1515
Size 3 counts
R
4
0–4,294,967,295
1517
Size 4 counts
R
4
0–4,294,967,295
1519–1542
Reserved
1543
Analog channel 1 (flow)
R
2
mV
1544
Analog channel 2 (temperature)
R
2
0.1 ºC external probe only
1545
Analog channel 3 (relative humidity)
R
2
0.1% RH external probe only
1546
Analog channel 4
R
2
1547
Analog channel 5 (CAL)
R
2
mV
1548–1552
Reserved
1553
Location name
R
32
Double byte characters (16)
1585-1599
Expansion
Содержание Met One 6000 Series
Страница 2: ......
Страница 33: ...31 Operation Figure 21 Setup utility program non wireless units...
Страница 48: ...46 Troubleshooting...
Страница 50: ...48 Replacement parts and accessories...
Страница 52: ...50 Contact information...
Страница 54: ...52 Limited warranty...
Страница 64: ...62 Modbus register maps...