33
Operation
Met One 6000 counters with the RS485 Modbus communication option use
industry-standard Modbus RTU protocol. In this communication mode, a series of
registers hold data about measurement results and operation parameters. The
parameters are preset by the user through a utility setup program or through the central
monitoring software. A list of tables in
shows the register
addresses, types and use. A user must write drivers to communicate with the particle
counter through these registers with the Modbus RTU protocol.
Set the instrument address—RS485
Important note: Address 0 can only be used with FXB protocol. Address 0 is reserved for
use as a broadcast address for Modbus RTU.
For a network of counters with RS485 Modbus or FXB protocol, use the dip switch on the
bottom of the counter to set the address (refer to
Table 8 Dip switch settings for network address
Network address
Dip switch 1
Dip switch 2
Dip switch 3
Dip switch 4
Dip switch 5
0
1
Off
Off
Off
Off
Off
1
On
Off
Off
Off
Off
2
Off
On
Off
Off
Off
3
On
On
Off
Off
Off
4
Off
Off
On
Off
Off
5
On
Off
On
Off
Off
6
Off
On
On
Off
Off
7
On
On
On
Off
Off
8
Off
Off
Off
On
Off
9
On
Off
Off
On
Off
10
Off
On
Off
On
Off
11
On
On
Off
On
Off
12
Off
Off
On
On
Off
13
On
Off
On
On
Off
14
Off
On
On
On
Off
15
On
On
On
On
Off
16
Off
Off
Off
Off
On
17
On
Off
Off
Off
On
18
Off
On
Off
Off
On
19
On
On
Off
Off
On
20
Off
Off
On
Off
On
21
On
Off
On
Off
On
22
Off
On
On
Off
On
23
On
On
On
Off
On
24
Off
Off
Off
On
On
25
On
Off
Off
On
On
26
Off
On
Off
On
On
27
On
On
Off
On
On
28
Off
Off
On
On
On
29
On
Off
On
On
On
30
Off
On
On
On
On
31
On
On
On
On
On
Содержание Met One 6000 Series
Страница 2: ......
Страница 33: ...31 Operation Figure 21 Setup utility program non wireless units...
Страница 48: ...46 Troubleshooting...
Страница 50: ...48 Replacement parts and accessories...
Страница 52: ...50 Contact information...
Страница 54: ...52 Limited warranty...
Страница 64: ...62 Modbus register maps...