4
Specifications
Counting efficiency
Model 6003: 50% (± 20%) for 0.3 µm, (100% ± 10% at 1.5 times the minimum
sensitivity). Fully complies with ISO21501-4.
Model 6005: 50% (± 20%) for 0.5 µm, (100% ± 10% at 1.5 times the minimum
sensitivity). Fully complies with ISO21501-4.
Model 6015: 50% (± 20%) for 0.5 µm, (100% ± 10% at 1.5 times the minimum
sensitivity). Fully complies with ISO21501-4.
Coincidence loss
Model 6003/6005 (all output options): 5% at 70,600,000 particles/m
3
(2,000,000 particles/ft
3
)
Model 6015 (all output options except for pulse): 5% at 14,000,000 particles/m
3
(400,000 particles/ft
3
)
False count rate
One or less in five minutes
Wireless
Network standards
IEEE 802.11b; IEEE 802.11g
Frequency range
2.412–2.484 GHz
Antenna connector
1, no diversity supported. Impedance 50 ohms
Data rates
1, 2, 5.5, 11 Mbps (802.11b); 6, 9, 12, 18, 24, 36, 48, 54 Mbps (802.11g)
Number of selectable sub-channels
Up to 14 channels. Profiles available will include USA, France, Japan, Spain,
Canada and "Other" (multiple countries)
Security
WEP 64/128, WPA, WPA2, PSK, TKIP
Range
Up to 91m (300 ft) indoors
1
Transmit output power
14 dBm or 25 milliwatts
Protocols supported
ARP, UDP, TCP, DHCP, Auto IP
Power consumption, maximum
9.7 W (wireless unit)
Certifications
Instrument
CE mark
2
Power Supply PULS CS5 Series
Certified to UL & CSA standards by UL (cULus mark) and CE mark
2
1
Maximum range may be reduced depending on several factors. Refer to
2
Refer to the product Declaration of Conformance (available on request).
Figure 1 Met One 6000 dimensions
Содержание Met One 6000 Series
Страница 2: ......
Страница 33: ...31 Operation Figure 21 Setup utility program non wireless units...
Страница 48: ...46 Troubleshooting...
Страница 50: ...48 Replacement parts and accessories...
Страница 52: ...50 Contact information...
Страница 54: ...52 Limited warranty...
Страница 64: ...62 Modbus register maps...