69
Standard Operations
requires a new baseline scan. When the baseline has been
scanned, the instrument is ready to scan one or more samples.
1.
Press
Wavelength Scan
in the Main Menu.
2.
Insert the blank cuvette/cell into the cuvette/sample cell holder.
3.
Press
Zero
. ”Zeroing“ appears below the graph as the baseline
scan begins.
4.
Insert the prepared sample cuvette/cell into the cuvette/sample
cell holder.
5.
Press
Read
. Below the graph ”Reading“ appears and a graph
of the absorbance or transmittance values at the scanned
wavelengths is displayed continuously.
The Wavelength Scan is complete, if
•
the graph is shown full size,
•
the scaling of the x-axis fits automatically,
•
the Cursor functions in the vertical navigation bar are
highlighted.
The navigation functions of the wavelength scan graph or a wavelength scan analysis are described in
Table 10 Navigating the wavelength scan
Cursor Function/
Zoom Function
Description
Curve Icon
(Choice of Cursor Mode)
Choice of Cursor Mode
Peak/Valley
(cursor moves between minimum/maximum absorbance
values) or Cursor Mode
Tracking
(cursor moves over each data point of the scan).