Standard Operations
5.7.2.2 Complete a measurement in the multi wavelength mode
1.
Insert the blank cuvette/cell into the cuvette/sample cell holder.
Press
Zero
.
Note:
The
Read
key does not become active until the zero
measurement has been completed.
2.
Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press
Read
.
3.
For data storage, see
.
5.7.3 Wavelength scan mode – recording of absorbance and transmission spectrums
In the wavelength scan mode, the absorbance of the light in a
solution over a defined wavelength spectrum is measured.
The measurement results can be displayed as a curve, as
percentage transmittance (%T) or as Absorbance (Abs). The
collected data can be printed as a table or a curve.
The data are available for formatting changes. These include
automatic scaling and zoom functions. Maximum and minimum
values are determined and shown as a table.
The cursor can be moved to any point on the curve for the purpose
of reading off the absorbance or transmittance value and the
wavelength. The data associated with each data point can also be
shown as a table.