W A R N I N G
Fire hazard. Battery substitution is not permitted. Use only alkaline
batteries.
N O T I C E
Make sure to tighten the screws to 11 N·m (8 ft·lb) for a correct seal fit and to
maintain the enclosure environmental rating.
Refer to
on page 6 for battery replacement.
Troubleshooting
Error
code
Possible cause
Solution
1
Instrument is not configured.
Contact technical support.
2
Could not read program data
3
Could not write program
data
4
Battery error
Replace the batteries.
5
A/D error during a
measurement
Contact technical support.
6
Offset error during a
measurement
Make sure that the instrument cap is
correctly installed.
Error
code
Possible cause
Solution
7
Low light error during a
measurement
• Make sure that there is no light
path blockage.
• Zero is out of instrument range
• Contact technical support.
8
Over range error during a
measurement, a potential
light blockage or the
concentration is too high
• Make sure that there is no light
path blockage and that the
concentration is not too high for
the program.
• Make sure that the instrument cap
is correctly installed.
• Contact technical support.
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment malfunction.
The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Replacement parts
Description
Item no.
Adapter, assembly, COD
4846400
Battery set, AA alkaline batteries (4x)
1938004
Sample cell with cap, 25 x 95 mm, 10-20-25 mL (6x)
2401906
Sample cell with cap, 1 cm/10 mL (2x)
4864302
USB cable
LZV818
16
English