Figure 3 Instrument components
1
Instrument
4
Power cord
7
Printer paper (2x)
2
AC power adapter
5
Sample bottle (10x)
8
Cleaning brush (2x)
3
CO
2
bottle, empty
(2x)
6
Allen wrench
9
Software disc
Installation
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Connect external pressure
Connect the supplied CO
2
bottle or shop air to the instrument to supply
external pressure.
Connect the CO
2
bottle
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
D A N G E R
Explosion hazard. For the safe use of the instrument, obey the
precautions and the instructions as specifically described.
W A R N I N G
Personal injury hazard. Enclosed systems contain high pressure.
Qualified personnel must remove pressure from the system before the
instrument can be installed or removed.
C A U T I O N
Fire hazard. Do not use flammable agents to clean the instrument.
The CO
2
bottle is a compressed gas bottle under high pressure. Do not
alter this bottle in any way. Use only CO
2
bottles that are supplied by the
manufacturer. Use the bottle only as directed in this manual. For storage
and disposal of the bottle, refer to the text on the bottle.
Do not expose the pressurized bottle to temperatures in excess of
177 °C (350 °F). Bottles exposed to a fire or heated to temperatures in
excess of 177 °C (350 °F) should be discarded. Let the temperature of
the bottle cool before it is discarded.
Keep the protective cap on the CO
2
bottles that are not in use.
Valves must be installed or removed only by trained personnel.
Use only a mild detergent and water to clean the cylinder. Use of caustic
cleaners may corrode pressure fittings or tank walls and cause an
uncontrolled pressure release or rupture.