See the system status menu parameters
Push
F3
(SYS) in the main menu to get the system status menu.
shows various system parameters that the instrument measures. These
system parameters show convenient information and diagnostic
troubleshooting.
Table 1 System parameters
Parameter
Description
Power supply
Shows whether the instrument is connected to the external
power supply or it is operated from the internal battery. When
an external power supply is used, the battery icon shows that
the internal battery is controlled by the external power supply.
When the internal battery is used, the battery icon shows the
remaining relative battery supply voltage capacity. The useful
battery voltage range for the instrument is 11.5 to 13.8 volts.
The amount of battery life that stays below this range is
minimal, so charge the battery before operation is continued.
Internal
temperature
Shows only the temperature inside the instrument enclosure.
This measurement tells if the temperature is too hot or cold to
do a sample test.
Printer status
Shows the different status conditions or errors of the internal
printer. If the printer is good, a condition of NORMAL shows.
An error message shows if a problem exists.
For an explanation of these error codes, refer to
on page 19.
Set the operation variables
The operation variables are located under the Setup menu. Refer to
. To set an operation variable:
1.
Push
F1
(SETUP) to get the Setup menu.
2.
Push the numeric number on the keypad that corresponds to the
number on the left side of the variable to program.
3.
Push
F1
(PRG) and use the navigation keys to select the proper
setting or use the keypad to input numbers, letters or symbols.
To add letters:
a.
Push the numeric key that contains the letter to use in
succession.
b.
Push once for the number, twice for the first letter, three times for
the second letter, etc. The time between the key pushes tells if
the current character is selected or if the next character is
selected.
To add symbols:
a.
Push the function key that shows the symbol to be used.
b.
Push
F1
for a space,
F2
for a forward slash and
F3
for a period.
4.
Push
F4
to accept the new setting or push
CANCEL
to return to the
Setup menu and keep the previous selection.
5.
To see the test results in a different Standard, Viscosity or
Temperature, set the variables and see or print the test results from
the buffer. Refer to
on page 15.
10
English