23
Installation
4.
Tighten the clamp and measure the frame position again to make sure it is at the
correct position.
3.2.2.6 Align the sensor horizontally
The sensor must be aligned horizontally to make sure that the sensor is centered over
the flow. If the pipe is not level and has a slope of 2 degrees or more, align the sensor to
be parallel with the surface of the water.
Prerequisites
•
Bubble level
Procedure
1.
Remove the paper backing from the bubble level and attach the level to the sensor
(
Figure 10 on page 21
).
2.
Loosen the clamps and tap the frame into position.
3.
Tighten both clamps and measure the frame position to make sure it is at the correct
position.
Figure 12 Vertical alignment of Flo-Dar sensor with SVS
1
Spacer
3
SVS sensor (optional)
2
Distance from crown of pipe to top of frame