visit to change the batteries and prevent lost data. If the logger contains a port for a rain
gauge, the Rain channel is also shown. To include a Logger channel in the datalog, select the
check box next to the channel name.
b.
Expand the tree for each Port[1](Sensor Name) channel group to view the available channels
for the sensor. If the check box next to Logger or Port[1] is selected, all of the channels in the
group are automatically selected.
c.
To include a Sensor channel in the datalog, select the check box next to the channel name.
The log channel count increases each time a channel is selected.
Note: For loggers with multiple sensor ports, the port number is added to the channel name. For example,
Velocity 3 is the velocity channel name for sensor port 3.
4.
To set the logging interval for a channel group:
a.
Click on the channel group name, for example
Port1 (Flo-Dar)
. The logging intervals are
shown.
b.
Select the interval from the drop-down list. The primary logging interval is used for normal
operation. The secondary logging interval is used during alarm conditions.
Note: The logging interval cannot be set for an individual sensor channel.
5.
Click
WRITE TO LOGGER
to save the settings. A message window is shown:
Option
Description
Warning: all data
will be lost.
Continue?
All data that is stored in the logger is erased when a program is written to the logger.
To save the data, select No and download the data to a safe location. Select Yes to
erase all data and update the logger with the new program.
Set Logger Clock
Synchronize to Computer Clock—the logger uses the date and time settings of the
computer. Set Logger Clock—the logger uses the date and time settings that are set
by the user. If the unit has a modem, the logger automatically uses the date time
settings of the server.
Calibrate the sensor with the Cal Wizard
Pre-requisite:
The logger must be connected to the computer and must be online for calibration.
The sensor can be configured and calibrated with the calibration wizard.
1.
Click the Sensors tab.
2.
Click on Sensor Port[1] (sensor name).
3.
Click on the
CAL WIZARD
button. The Calibration Wizard window opens.
English
21