64
Advanced Operations
Selecting the Concentration Resolution:
1.
From the Options menu, press
CONCENTRATION
RESOLUTION
.
2.
Select the resolution and press
OK
.
6.4.2 Performing a Measurement in the Multi-Wavelength Mode
1.
Insert the blank cell into the cell holder. Press
ZERO
.
Note: The
READ
key does not become active until the zero
measurement has been performed.
2.
Insert the sample cell into the cell holder. Press
READ
.
3.
For data storage, see
section 5.3.1 on page 28
.
6.5 Time Course Mode
The Time Course Mode is used to collect data in either absorbance
or transmittance for a user-specified length of time. After the data
are collected, they can be displayed in either graphic or tabular
format.
6.5.1 Time Course Setup Parameters
1.
From the Main Menu , press
TIME COURSE
.
2.
Press
OPTIONS
to configure parameters.