Hach CDC401 Series Скачать руководство пользователя страница 9

• The meter can show the measured value as conductivity, TDS (total dissolved solids), salinity or

resistivity. Use the probe settings menu to change the parameter.

• If the probe is used to measure the salinity value for dissolved oxygen measurements, change the

parameter to salinity.

• The conductivity value changes with temperature. The default probe settings apply a temperature

correction to adjust the conductivity measurements to the value at 25 °C.

• The meter automatically saves the measurement data when the user manually reads each data

point and when the meter is set to read at regular intervals. The user must manually save each
data point when the meter is set to read continuously.

• Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure

to remove air bubbles before and during measurements.

• If the probe is a rugged type, make sure to install the shroud before field use to prevent damage to

the sensing elements. Refer to 

Remove or install the shroud

 on page 9. The probe warranty

does not include such damage.

• To deploy a rugged probe at a distance, toss the probe body with a slow underhand throw. Do not

throw the probe by the cable to prevent damage to cable, the probe or the user.

6.2 Sample measurement procedure

1. 

 Rinse the probe

with deionized water.
Dry the probe with a
lint-free cloth.
Rugged probes:
install the shroud.

2. 

 Put the probe in

the sample with the
sensor fully in the
sample. Do not put
the probe on the
bottom or sides of
the beaker.

3. 

 Shake the probe

from side to side to
remove air bubbles.
Rugged probes:
move the probe up
and down to remove
air bubbles.

4. 

 Stir gently, then

read the conductivity
(or TDS, salinity,
resistivity) value of
the sample. The
display shows the
temperature-
corrected value
when the reading is
stable.

Section 7 Verify the calibration

Measure the value of a fresh conductivity standard solution to make sure the result is accurate. The
meter compares the selected standard solution value to the measured value and accepts or rejects
the measurement. The user can change the standard solution and acceptance criteria for verification
in the probe-specific settings.

Note: Password protection may prevent access to the acceptance criteria.

English

   

7

Содержание CDC401 Series

Страница 1: ...DOC022 53 80022 CDC401 05 2021 Edition 4 User Manual...

Страница 2: ......

Страница 3: ...4 Preparation for use 5 Section 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 6 6 2 Sample measurement procedure 7 Sec...

Страница 4: ...Table of Contents 2...

Страница 5: ...uctivity cell 3 Shroud rugged model 2 Standard probe 4 Rugged probe Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Graphite 4 pole condu...

Страница 6: ...0 m 98 4 ft Warranty 1 year on the probe This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical...

Страница 7: ...the meter Note Some meters automatically open the date and time settings when the meter starts for the first time or after battery replacement 3 Connect the probe to the meter Section 5 Calibration T...

Страница 8: ...put the probe on the bottom or sides of the beaker 5 Shake the probe from side to side to remove air bubbles 6 Stir gently then read the conductivity value of the standard solution The display shows t...

Страница 9: ...warranty does not include such damage To deploy a rugged probe at a distance toss the probe body with a slow underhand throw Do not throw the probe by the cable to prevent damage to cable the probe o...

Страница 10: ...ndard solution The meter accepts or rejects the result Section 8 Maintenance 8 1 Clean the probe Clean the probe regularly to remove mineral or sample buildup on the electrodes Symptoms of contaminati...

Страница 11: ...oils and grease contamination KATHON CG TRITON X 2 hours maximum Proteins Electrode cleaning solution for proteins organic contamination Pepsin in HCl 3 hours maximum Wastewater and organic compounds...

Страница 12: ...asurements The sample absorbs carbon dioxide CO2 from the air which forms carbonic acid and slowly increases the conductivity in low ionic strength LIS or high purity samples Use the LIS chamber for L...

Страница 13: ...0 pkg 2771320 Conductivity standard solution KCl 1413 S cm Singlet one use packets 20 mL each 20 pkg 2771420 Conductivity standard solution KCl 12 88 mS cm Singlet one use packets 20 mL each 20 pkg 27...

Страница 14: ...12 English...

Страница 15: ......

Страница 16: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: