Table 1 Cleaning solution
Contamination
Cleaning solution
Active component
Soak time
General
contamination
Electrode cleaning solution for
regular maintenance
KATHON
™
CG,
DECONEX
®
11
12–16 hours
Minerals
Electrode cleaning solution for
minerals/inorganic
contamination
Phosphoric acid
(~10%)
10–15 minutes
Fats, grease and oils
Electrode cleaning solution for
fats, oils and grease
contamination
KATHON
™
CG,
TRITON
®
X
2 hours maximum
Proteins
Electrode cleaning solution for
proteins/organic
contamination
Pepsin in HCl
3 hours maximum
Wastewater and
organic compounds
Electrode cleaning solution,
extra strong
Sodium hypochlorite
5–10 minutes
8.2 Remove or install the shroud
Remove the shroud on the rugged probe during calibration and maintenance. Refer to
Keep the shroud installed on the rugged probe during sample measurements to prevent damage to
the sensor. Refer to
Figure 2 Remove the shroud