Table 1  Cleaning solution

Contamination

Cleaning solution

Active component

Soak time

General
contamination

Electrode cleaning solution for
regular maintenance

KATHON

 CG,

DECONEX

®

11 

12–16 hours

Minerals

Electrode cleaning solution for
minerals/inorganic
contamination

Phosphoric acid
(~10%)

10–15 minutes

Fats, grease and oils

Electrode cleaning solution for
fats, oils and grease
contamination

KATHON

 CG,

TRITON

®

 X

2 hours maximum

Proteins

Electrode cleaning solution for
proteins/organic
contamination

Pepsin in HCl

3 hours maximum

Wastewater and
organic compounds

Electrode cleaning solution,
extra strong

Sodium hypochlorite

5–10 minutes

8.2 Remove or install the shroud

Remove the shroud on the rugged probe during calibration and maintenance. Refer to 

Figure 2

.

Keep the shroud installed on the rugged probe during sample measurements to prevent damage to
the sensor. Refer to 

Figure 3

.

Figure 2  Remove the shroud

English

   

9

Содержание CDC401 Series

Страница 1: ...DOC022 53 80022 CDC401 05 2021 Edition 4 User Manual...

Страница 2: ......

Страница 3: ...4 Preparation for use 5 Section 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 6 6 2 Sample measurement procedure 7 Sec...

Страница 4: ...Table of Contents 2...

Страница 5: ...uctivity cell 3 Shroud rugged model 2 Standard probe 4 Rugged probe Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Graphite 4 pole condu...

Страница 6: ...0 m 98 4 ft Warranty 1 year on the probe This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical...

Страница 7: ...the meter Note Some meters automatically open the date and time settings when the meter starts for the first time or after battery replacement 3 Connect the probe to the meter Section 5 Calibration T...

Страница 8: ...put the probe on the bottom or sides of the beaker 5 Shake the probe from side to side to remove air bubbles 6 Stir gently then read the conductivity value of the standard solution The display shows t...

Страница 9: ...warranty does not include such damage To deploy a rugged probe at a distance toss the probe body with a slow underhand throw Do not throw the probe by the cable to prevent damage to cable the probe o...

Страница 10: ...ndard solution The meter accepts or rejects the result Section 8 Maintenance 8 1 Clean the probe Clean the probe regularly to remove mineral or sample buildup on the electrodes Symptoms of contaminati...

Страница 11: ...oils and grease contamination KATHON CG TRITON X 2 hours maximum Proteins Electrode cleaning solution for proteins organic contamination Pepsin in HCl 3 hours maximum Wastewater and organic compounds...

Страница 12: ...asurements The sample absorbs carbon dioxide CO2 from the air which forms carbonic acid and slowly increases the conductivity in low ionic strength LIS or high purity samples Use the LIS chamber for L...

Страница 13: ...0 pkg 2771320 Conductivity standard solution KCl 1413 S cm Singlet one use packets 20 mL each 20 pkg 2771420 Conductivity standard solution KCl 12 88 mS cm Singlet one use packets 20 mL each 20 pkg 27...

Страница 14: ...12 English...

Страница 15: ......

Страница 16: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: