Figure 3  Install the shroud

8.3 Storage

Between uses, make sure the probe is dry and store it in ambient conditions. Rugged probes may be
stored with the shroud installed if the storage container is sufficiently large.

Section 9 Troubleshooting

Problem

Possible cause

Solution

Decreased probe performance
causes slow stabilization and
prevents accurate calibrations
or measurements.

The probe has mineral or sample
buildup on the electrodes.

Clean the probe. Refer to 

Clean the probe

 on page 8.

Sample properties cause slow
stabilization or inaccurate
measurements.

The sample absorbs carbon
dioxide (CO

2

) from the air, which

forms carbonic acid and slowly
increases the conductivity in low
ionic strength (LIS) or high purity
samples.

Use the LIS chamber for
LIS/high purity samples to
prevent CO

2

 absorption.

Procedure problem causes
slow stabilization and prevents
accurate calibrations or
measurements.

An incorrect standard solution was
used or the standard solution is
contaminated.

Use the specified standard
solution of good quality.

Air bubbles are around or below
the probe tip.

Carefully tap or shake the
probe to remove air bubbles.

The conductivity settings in the
meter are not correct.

Make sure that the settings
for temperature correction,
reference temperature and
stability are correct.

10

   

English

Содержание CDC401 Series

Страница 1: ...DOC022 53 80022 CDC401 05 2021 Edition 4 User Manual...

Страница 2: ......

Страница 3: ...4 Preparation for use 5 Section 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 6 6 2 Sample measurement procedure 7 Sec...

Страница 4: ...Table of Contents 2...

Страница 5: ...uctivity cell 3 Shroud rugged model 2 Standard probe 4 Rugged probe Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Graphite 4 pole condu...

Страница 6: ...0 m 98 4 ft Warranty 1 year on the probe This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical...

Страница 7: ...the meter Note Some meters automatically open the date and time settings when the meter starts for the first time or after battery replacement 3 Connect the probe to the meter Section 5 Calibration T...

Страница 8: ...put the probe on the bottom or sides of the beaker 5 Shake the probe from side to side to remove air bubbles 6 Stir gently then read the conductivity value of the standard solution The display shows t...

Страница 9: ...warranty does not include such damage To deploy a rugged probe at a distance toss the probe body with a slow underhand throw Do not throw the probe by the cable to prevent damage to cable the probe o...

Страница 10: ...ndard solution The meter accepts or rejects the result Section 8 Maintenance 8 1 Clean the probe Clean the probe regularly to remove mineral or sample buildup on the electrodes Symptoms of contaminati...

Страница 11: ...oils and grease contamination KATHON CG TRITON X 2 hours maximum Proteins Electrode cleaning solution for proteins organic contamination Pepsin in HCl 3 hours maximum Wastewater and organic compounds...

Страница 12: ...asurements The sample absorbs carbon dioxide CO2 from the air which forms carbonic acid and slowly increases the conductivity in low ionic strength LIS or high purity samples Use the LIS chamber for L...

Страница 13: ...0 pkg 2771320 Conductivity standard solution KCl 1413 S cm Singlet one use packets 20 mL each 20 pkg 2771420 Conductivity standard solution KCl 12 88 mS cm Singlet one use packets 20 mL each 20 pkg 27...

Страница 14: ...12 English...

Страница 15: ......

Страница 16: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: