13. 

 Push and hold

ON

 to start a test. A

graph will show.

Note: To cancel a test,
push and hold OFF.
There is a built-in 1
hour instrument/sample
equilibration period
before data collection.
The display shows
"DELAY" during this
period.

14. 

 Do steps 10

through 13 again to
set the test range
and start each of the
six channels. It is not
necessary to operate
all six channels if
less than six
samples are
available.

Section 8 Determination of results

After the end of the selected test period (5, 7 or 10 days), the display
shows "END". The results are given based on the selected procedure:
Simplified or Hach Standard Method.

8.1 Simplified sample results

The simplified procedure results are shown on the BODTrak II display.
Push the applicable channel selection key to show the results.

Note: If the sample was pre-diluted, apply a dilution factor to the instrument
reading. Refer to 

Sample dilution

 on page 15.

8.2 Hach Standard Method results

1.

Push the channel selection key for the Hach Standard Method
sample bottles, and the seed blank bottle and GGA bottle when
applicable. The results are shown.

Note: Treat the seed blank the same as all other samples. If the sample was
pre-diluted, apply a dilution factor to the instrument reading (

Sample dilution

on page 15).

2.

Find the dilution factor based on the selected range (

Table 5

on page 11). For example, if the sample range selected is 0 to
350 mg/L BOD, the dilution factor is 1.45.

3.

Calculate the corrected results:
BOD mg/L = BOD mg/L (instrument reading) x dilution factor

Example:

• Instrument Reading = 200 mg/L
• BOD dilution factor = 1.45 

200 mg/L x 1.45 = 290 mg/L BOD (corrected result)

4.

When the samples are seeded (not GGA), calculate the results with
the equation that follows and the corrected results:
BOD (mg/L) = A – [B x (SA÷SB)]
Where:

• A = corrected BOD of the seeded sample
• B = corrected BOD of the seed blank
• SA = volume of seed in seeded sample
• SB = volume of sample in seeded sample

Example:

• A = 290 mg/L BOD
• B = 120 mg/L BOD
• SA = 20 mL
• SB = 110 mL

BOD (mg/L) = 290 mg/L - [120 mg/L x (20 mL÷110 mL)]
BOD mg/L = 268 mg/L

5.

For GGA, calculate the results of the GGA with the equation that
follows and the corrected BOD of the seed blank:
BOD (mg/L) = C – [B x (SA÷SC)]
Where:

• C = instrument reading on seeded GGA

14

   

English

Содержание BODTrak II

Страница 1: ...DOC022 53 80363 BODTrak II 11 2020 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...tions 5 3 2 Connect the RS232 interface 6 3 3 Bottle connections 6 Section 4 User interface 6 4 1 Channel selection keys 7 4 2 Arrow keys 7 4 3 ON key 7 4 4 OFF key 7 Section 5 Startup 7 5 1 Turn the...

Страница 4: ...10 4 Toxic materials 16 10 5 Chlorine 16 10 6 pH effect 16 10 7 Supersaturation 16 Section 11 Maintenance 17 11 1 Cleaning 17 11 2 Storage 17 Section 12 Troubleshooting 17 12 1 Incorrect BOD curves 17...

Страница 5: ...rd 150 mg L each of glucose and glutamic acid Number of samples 44 Number of analysts 1 Number of BodTrak II instruments 6 Results Mean of 235 mg L BOD Distribution 11 mg L or range of 224 to 246 mg L...

Страница 6: ...references the instruction manual for operation and or safety information Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems Return ol...

Страница 7: ...es is monitored Bacteria in the sample use oxygen when they consume organic matter This oxygen consumption causes the pressure in the bottle head space to drop The pressure drop correlates directly to...

Страница 8: ...s if the device cannot continuously receive at 9600 baud Note The use of the specified cable or an equivalent shielded cable is necessary to meet radio frequency emissions requirements 3 3 Bottle conn...

Страница 9: ...test length and range 4 3 ON key To go to the range selection menu push ON from a channel display screen Then push and hold ON to start the test for the selected channel 4 4 OFF key When a test is in...

Страница 10: ...zard select Delimited as the file type the start row in the spreadsheet and Windows ANSI as the file origin Click NEXT 4 Click the check boxes for Space delimiter and Treat consecutive delimiters as o...

Страница 11: ...propriate to the chemicals that are handled Refer to the current safety data sheets MSDS SDS for safety protocols C A U T I O N Chemical exposure hazard Dispose of chemicals and wastes in accordance w...

Страница 12: ...level the 0 to 35 mg L BOD range is decreased to 0 to 25 mg L BOD Adjustment is not necessary for other test ranges Refer to Special considerations on page 15 for special considerations including samp...

Страница 13: ...e size If samples are seeded prepare a seed blank before a sample is prepared Treat the seed blank the same as any other sample and omit step 5 Use Hach BOD Standard Solution Ampules for Manometric Me...

Страница 14: ...eded use a tensette pipet to add the correct quantity of seed to the graduated cylinder Table 5 6 If necessary add more nutrient or buffer Do not add more than a total volume of 50 mL seed nutrient bu...

Страница 15: ...s are rotating If not lift the bottle up then set it down again 7 Push and hold the LEFT and RIGHT arrows at the same time to go to the instrument setup menu Note Set the time and date if necessary Se...

Страница 16: ...s and the seed blank bottle and GGA bottle when applicable The results are shown Note Treat the seed blank the same as all other samples If the sample was pre diluted apply a dilution factor to the in...

Страница 17: ...r to the expanded version of the manual for information about specific procedures Figure 3 shows typical curves through a 10 day test period For incorrect curves refer to Figure 4 on page 18 Figure 3...

Страница 18: ...4 Toxic materials Industrial and chlorinated samples often contain toxic substances and require special considerations when running BOD tests Toxic materials in the sample will cause decreased BOD va...

Страница 19: ...onized or distilled water Clean the sample bottles Clean the sample bottles and caps with a brush water and a mild detergent Flush the sample bottles with fresh water followed by a distilled water rin...

Страница 20: ...fluent is in the 0 70 mg L range while influent is in the 0 700 mg L range when the BOD of the sample is more than 700 mg L prepare a sample dilution Refer to the Sample dilution section in the expand...

Страница 21: ...leak makes the system unresponsive If such a condition occurs examine the seal cup and bottle cap for contamination or damage Section 13 Replacement parts and accessories 13 1 Replacement parts Descr...

Страница 22: ...size 2 1 68700 Buret straight stopcock 25 mL 1 1405940 Clamp buret double 1 32800 Cylinder graduated 10 mL 1 50838 Cylinder graduated 25 mL 1 50840 Cylinder graduated 50 mL 1 50841 Cylinder graduated...

Страница 23: ......

Страница 24: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1...

Отзывы: