Use the applicable sample volume tables for each procedure.

If power is interrupted when the instrument is in 

DELAY

 status, the

test will stop and the status will change to 

CLEAR

 when power

returns. Start the test again. If power is interrupted when the
instrument is in 

RUN

 status, the test will resume when power returns.

Keep deionized water overnight in an incubator at 20 ºC. Shake the
deionized water to saturate with air.

Let the seed become stable overnight in the BOD incubator at a
temperature of 20 ºC. Be careful not to disturb the settled solution.
Pipet seed solution from the top.

Dilution is necessary if samples have BOD values more than
700 mg/L.

At elevations higher than 5000 feet above sea level, the 0 to 35 mg/L
BOD range is decreased to 0 to 25 mg/L BOD. Adjustment is not
necessary for other test ranges.

Refer to 

Special considerations

 on page 15 for special

considerations including sample seeding and pretreatment.

Use only BODTrak II stir bars and bottles. These items are made
specifically to be used with the BODTrak II.

7.2 Simplified procedure

Items to collect:

• BODTrak II bottle
• Thermometer
• Blender (optional)
• Graduated cylinder
• Nutrient buffer pillow (1x)

Use the correct sample volume for the BOD range as shown in 

Table 4

.

Table 4  Simplified sample volumes

BOD range (mg/L)

Sample volume (mL)

0 to 35 

420 

0 to 70 

355 

0 to 350 

160 

0 to 700 

95 

1. 

 Heat or cool the

sample to 19 to
21 ºC.

2. 

 Homogenize the

sample in a blender
if it contains large
settleable or
floatable solids.

3. 

 Select the correct

sample volume for
the BOD range
(

Table 4

). Measure

the sample into a
graduated cylinder.

10

   

English

Содержание BODTrak II

Страница 1: ...DOC022 53 80363 BODTrak II 11 2020 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...tions 5 3 2 Connect the RS232 interface 6 3 3 Bottle connections 6 Section 4 User interface 6 4 1 Channel selection keys 7 4 2 Arrow keys 7 4 3 ON key 7 4 4 OFF key 7 Section 5 Startup 7 5 1 Turn the...

Страница 4: ...10 4 Toxic materials 16 10 5 Chlorine 16 10 6 pH effect 16 10 7 Supersaturation 16 Section 11 Maintenance 17 11 1 Cleaning 17 11 2 Storage 17 Section 12 Troubleshooting 17 12 1 Incorrect BOD curves 17...

Страница 5: ...rd 150 mg L each of glucose and glutamic acid Number of samples 44 Number of analysts 1 Number of BodTrak II instruments 6 Results Mean of 235 mg L BOD Distribution 11 mg L or range of 224 to 246 mg L...

Страница 6: ...references the instruction manual for operation and or safety information Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems Return ol...

Страница 7: ...es is monitored Bacteria in the sample use oxygen when they consume organic matter This oxygen consumption causes the pressure in the bottle head space to drop The pressure drop correlates directly to...

Страница 8: ...s if the device cannot continuously receive at 9600 baud Note The use of the specified cable or an equivalent shielded cable is necessary to meet radio frequency emissions requirements 3 3 Bottle conn...

Страница 9: ...test length and range 4 3 ON key To go to the range selection menu push ON from a channel display screen Then push and hold ON to start the test for the selected channel 4 4 OFF key When a test is in...

Страница 10: ...zard select Delimited as the file type the start row in the spreadsheet and Windows ANSI as the file origin Click NEXT 4 Click the check boxes for Space delimiter and Treat consecutive delimiters as o...

Страница 11: ...propriate to the chemicals that are handled Refer to the current safety data sheets MSDS SDS for safety protocols C A U T I O N Chemical exposure hazard Dispose of chemicals and wastes in accordance w...

Страница 12: ...level the 0 to 35 mg L BOD range is decreased to 0 to 25 mg L BOD Adjustment is not necessary for other test ranges Refer to Special considerations on page 15 for special considerations including samp...

Страница 13: ...e size If samples are seeded prepare a seed blank before a sample is prepared Treat the seed blank the same as any other sample and omit step 5 Use Hach BOD Standard Solution Ampules for Manometric Me...

Страница 14: ...eded use a tensette pipet to add the correct quantity of seed to the graduated cylinder Table 5 6 If necessary add more nutrient or buffer Do not add more than a total volume of 50 mL seed nutrient bu...

Страница 15: ...s are rotating If not lift the bottle up then set it down again 7 Push and hold the LEFT and RIGHT arrows at the same time to go to the instrument setup menu Note Set the time and date if necessary Se...

Страница 16: ...s and the seed blank bottle and GGA bottle when applicable The results are shown Note Treat the seed blank the same as all other samples If the sample was pre diluted apply a dilution factor to the in...

Страница 17: ...r to the expanded version of the manual for information about specific procedures Figure 3 shows typical curves through a 10 day test period For incorrect curves refer to Figure 4 on page 18 Figure 3...

Страница 18: ...4 Toxic materials Industrial and chlorinated samples often contain toxic substances and require special considerations when running BOD tests Toxic materials in the sample will cause decreased BOD va...

Страница 19: ...onized or distilled water Clean the sample bottles Clean the sample bottles and caps with a brush water and a mild detergent Flush the sample bottles with fresh water followed by a distilled water rin...

Страница 20: ...fluent is in the 0 70 mg L range while influent is in the 0 700 mg L range when the BOD of the sample is more than 700 mg L prepare a sample dilution Refer to the Sample dilution section in the expand...

Страница 21: ...leak makes the system unresponsive If such a condition occurs examine the seal cup and bottle cap for contamination or damage Section 13 Replacement parts and accessories 13 1 Replacement parts Descr...

Страница 22: ...size 2 1 68700 Buret straight stopcock 25 mL 1 1405940 Clamp buret double 1 32800 Cylinder graduated 10 mL 1 50838 Cylinder graduated 25 mL 1 50840 Cylinder graduated 50 mL 1 50841 Cylinder graduated...

Страница 23: ......

Страница 24: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1...

Отзывы: