4.1 Channel selection keys
Push the related channel selection key to show data for one of the six
bottles. The channel selection keys are also used in the instrument setup
menu to select a parameter to be edited. Refer to
on page 6
.
Table 3 Channel key setup parameters
Channel
Parameter
1
Year (0–99)
2
Month (1–12)
3
Day (1–31)
4
Hour (0–24)
5
Minute (0–59)
6
Test length (5, 7 or 10 days)
4.2 Arrow keys
The display shows a graph of BOD values on the vertical axis and time
in days on the horizontal axis. Push the
LEFT
and
RIGHT
arrows to
move the cursor along the BOD curve to show the approximate
coordinates (time, BOD) of the selected data point.
The time interval and BOD value of the data point are shown in the lower
right of the display. The cursor is automatically placed at the most
recently collected data point in a channel display.
Push and hold the
LEFT
and
RIGHT
arrows at the same time to go into
the instrument setup menu. The arrow keys are also used to change the
time, date, test length and range.
4.3 ON key
To go to the range selection menu, push
ON
from a channel display
screen. Then push and hold
ON
to start the test for the selected channel.
4.4 OFF key
When a test is in
DELAY
or
RUN
modes, push and hold
OFF
to
manually end the test. The instrument will show
END
. The
OFF
key is
also used to exit the instrument setup menu or range selection menu.
The changes made before the menu is exited are saved.
Section 5 Startup
5.1 Turn the instrument on
Note: The ON and OFF keys are used to start and stop a test. They do not power
the instrument on and off.
1.
Connect the power adaptor to the DC power connector (
on page 6).
2.
Toggle the power switch to set the instrument to on and off. (
on page 6).
5.2 Set the clock
All the channels must show
END
or
CLEAR
before the clock can be set.
1.
Push and hold the two arrow keys at the same time until the
instrument setup menu is shown.
2.
Push the applicable channel key to select the clock parameter to be
adjusted.
3.
Use the arrow keys to edit the selected parameter. Adjust each
parameter in the same manner.
4.
When all the time adjustments are complete, push
OFF
to save and
go back to the data display screen.
Section 6 Standard operation
6.1 Typical curves
Refer to the expanded version of the manual for information about
specific procedures.
English
7
Содержание BODTrak II
Страница 110: ...15 15 1 2 15 1 2 3 4 5 2 6 mg L 360 2 7 BODTrak II UL CSA 115 VAC NEMA 5 15P 110...
Страница 112: ...9600 3 3 1 6 1 2 112 4 2 2 1 4 2 5 3 1 4 1 2 112 3 3 1 0 99 2 1 12 3 1 31 1 112...
Страница 116: ...RS232 111 2 3 5 127 7 MSDS SDS 7 1 7 2 8 8 1 4 10 113 116...
Страница 117: ...4 1 4 20 C 2 5 3 8 2 4 117 350 mg L 160 mL 1 GGA Hatch Hatch 0 68 0 70 mg L 0 700 mg L 700 mg L 8 3 4 117 2 5 Hach 117...
Страница 156: ...2 1 CD 2 2 2 3 2 4 2 5 ICES 003 A 156...
Страница 159: ...9600 baud 3 3 1 6 1 2 159 4 2 2 1 4 2 5 3 ON OFF 1 4 1 2 159 3 3 1 0 99 2 1 12 1 ON OFF 159...
Страница 163: ...RS232 158 2 3 5 127 7 MSDS SDS 7 1 Cleaning 7 2 8 8 1 BOD 4 BOD 10 160 163...
Страница 167: ......