18. Cal Signal (if configured in the system).
Use this function to test the Calibration Signal output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”. See Section
19. Maint Signal (if configured in the system).
Use this function to test the Maintenance Signal output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device
is on, it will be marked with an “*”. See Section
20. Remote Standby (if configured in the system).
Use this function to test the Remote Standby output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”. See Section
21. Stop (if configured in the system).
Use this function to test the Stop output. To change the state of the
device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If the device is
on, it will be marked with an “*”. See Section
22. Note (if configured in the system).
Use this function to test the system Notification output. To change
the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER key
again. If the device is on, it will be marked with an “*”. See Section
23. Warning (if configured in the system).
Use these functions to test the Warning output. To change the
state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If
the device is on, it will be marked with an “*”. See Section
24. Fault (if configured in the system).
Use this function to test the Fault output. To change the state of the
device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If the device is
on, it will be marked with an “*”. See Section
25. Man Mode Trig (if configured in the system).
Use this function to test system Manual Mode Trigger
output. To change the state of the device, press the ENTER key, set the device to ON/OFF, and press
the ENTER key again. If the device is on, it will be marked with an “*”. See Section
26. 4-20mA Chng (if configured in the system).
Use this function to test system generic 4-20mA Change
output signal. To change the state of the device, press the ENTER key, set the device to ON/OFF, and
press the ENTER key again. If the device is on, it will be marked with an “*”. See Section
27-29. 4-20mA Chng 1-3 (if configured in the system).
Use this function to test system 4-20mA Change
output signal for stream specific channels from Channel 1 to Channel 3. To change the state of the
device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If the device is
on, it will be marked with an “*”. See Section
30. 4-20mA Read (if configured in the system).
Use this function to test system 4-20mA Read output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”.See Section
31. DW Fail (if configured in the system).
Use this function to test system DW FAIL (Drinking Water %
REMOVAL FAIL) output. To change the state of the device, press the ENTER key, set the device to
ON/OFF, and pr
ess the ENTER key again. If the device is on, it will be marked with an “*”.See Section
33.
Input/Output Status.
Input/Output Status is a link to Maintenance, Diagnostics, Input/Output Status
Содержание BioTector B3500s
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Страница 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...