![Hach BioTector B3500s Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/hach/biotector-b3500s/biotector-b3500s_user-manual_3203147136.webp)
Page 136
8.3.14
Password
OPERATION 0000
CALIBRATION 0000
DIAGNOSTICS 0000
COMMISSIONING 0000
SYSTEM CONFIGURATION 0000
▪
Password menu allows any number from 1 to 9999 to be set as
passwords for the Operation, Calibration, Diagnostics,
Commissioning and System Configuration menus (levels). When
the setting is 0000 (by default), the password is disabled.
▪
The higher level passwords can be used to access password
protected lower levels. For instance the DIAGNOSTICS
password can be used for access the OPERATION level. The
SYSTEM CONFIGURATION password can be used to access
the password protected OPERATION, CALIBRATION,
DIAGNOSTICS and COMMISSIONING levels.
8.3.15
Language
ENGLISH
DEUTSCH
FRANCAIS
▪
LANGUAGE menu allows the user to change the language of
the system if it is available in the software.
8.3.16
Hardware Configuration
▪
Hardware Configuration menus are for factory use only.
Содержание BioTector B3500s
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Страница 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...