ZERO WATER NO
▪
When ZERO WATER
(programmed as “NO” by default)
is
programmed as “YES”, BioTector takes DI Water
from the
available system port (which could be the ZERO WATER port or
MANUAL/CALIBRATION port, and if these ports are not
available SAMPLE 1 port), and uses DI Water as a sample
during the Zero Calibration and Zero Check cycles.
▪
When this function is activated, the relevant system messages
are also displayed in Install New Reagents menu. See section
pH ADJUST BASE 14 , 0 , 0
▪
The three pH ADJUST BASE parameters defines the number of
pulses the base reagent pump operates to inject base reagent
into the reactor at each available system operation range, which
are Range 1, Range 2 and Range 3.
▪
This function is used to neutralize the pH of the Mixer Reactor.
pH ADJUST TIME 60s
▪
pH ADJUST TIME determines the mixing time (60 seconds by
default) the Mixer Reactor operates for the pH neutralization of
the reactor.
SERVICE ZERO 20 , 5
▪
SERVICE ZERO specifies the number of zero calibration cycles
to be repeated after a defined number of reactions between
each zero calibration cycle. The first setting “20” is the number of
reactions between each zero calibration cycle, and the second
setting “5” is the total number of zero calibration cycles which will
be carried out.
▪
SERVICE ZERO cycle is activated by the SET SERVICE ZERO
function in Service menu.
ZERO ADJUST HISTORY
▪
Anytime the Zero Adjust value is changed, either manually by
the user or automatically by the system, the new Zero Adjust
value is stored in the Zero Adjust History archive with the time,
date, zero range and the zero type (calibration, check or manual)
information.
▪
The meaning of the codes used in the system for zero adjust are
as follows:
TOC/TC-ZC: Zero calibration result for TOC/TC.
TOC/TC-ZK: Zero check result for TOC/TC.
TOC/TC-ZM: Manually input Zero Adjust for TOC/TC.
Содержание BioTector B3500s
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Страница 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...