W A R N I N G
Electrocution hazard. High voltage wiring for the controller is conducted behind the high voltage barrier in the
controller enclosure. The barrier must remain in place except when installing modules, or when a qualified
installation technician is wiring for power, relays or analog and network cards.
N O T I C E
Potential instrument damage. Delicate internal electronic components can be damaged by static electricity,
resulting in degraded performance or eventual failure.
To install the module and connect the sensor, refer to the illustrated steps that follow and the wiring
table.
Be sure to connect all sensor ground/shield wires to the controller enclosure grounding screws.
Note: If the sensor cable is not long enough to reach the controller, an interconnect cable (model 150727,10000)
and junction box (model 08335=A=6000) should be used to extend the distance.
Table 1 Conductivity sensor wiring
Terminal
Description
Sensor
8398
831x and 8394
1
Inner electrode
White (white cable)
White
2
Inner electrode
Black (white cable)
—
3
Ground
—
Green
4
—
—
—
5
—
—
—
6
—
—
—
7
Temp -
Green
Black
8
—
—
—
9
—
—
—
10
Temp +
Yellow
Blue
11
Outer electrode
Brown (brown cable)
Red
12
Outer electrode
Black (brown cable)
—
In
Step 3
of the following sequence, set the switch as follows:
Switch Setting
Description
Setting 1
Contacting conductivity sensor
Setting 2
Inductive conductivity sensor
6
English
Содержание 8332
Страница 2: ...English 3 Français 22 Español 42 中文 64 2 ...
Страница 7: ...English 7 ...
Страница 8: ...8 English ...
Страница 26: ...26 Français ...
Страница 27: ...Français 27 ...
Страница 46: ...46 Español ...
Страница 47: ...Español 47 ...
Страница 68: ...68 中文 ...
Страница 69: ...中文 69 ...
Страница 81: ......